Mindray Anesthetic Gas (AG) Module P/N 115-043908-00 For Hospital

 

 

The AG module analyzes gas samples from the patient and calculates CO2, O2, N2O, and AA waveforms and related numerics that include airway respiratory rate and MAC (minimum alveolar concentration).

Measurement range:

HAL, ENF, ISO, SEV, DES: 0~30 %,

N2O: 0~100 %;

awRR: 2~100 rpm

Accuracy:

Full accuracy mode: N2O: 0~20%REL: ±2% ABS, 20~100%REL: ±3%

 

ABS; HAL,ENF,ISO: 0~1%REL: ±0.15% ABS, 1~5%REL: ±0.2% ABS, >5%REL, not specified; SEV: 0~1%REL: ±0.15% ABS, 1~5%REL: ±0.2% ABS, 5~8%REL: ±0.4% ABS, >8%REL, not specified;

DES:0~1%REL: ±0.15% ABS, 1~5%REL: ±0.2% ABS, 5~10%REL: ±0.4% ABS, 10~15%REL: ±0.6% ABS ,15~18%REL: ±1% ABS ,>18%REL, not specified; awRR:2~60rpm, ±1rpm, >60rpm, not specified

This measurement can be used for adults, pediatrics and neonates.

 

 

Anesthetic Gas (AG) Module Alarm Messages

 

 

MESSAGE CAUSE

ALARM

PRIORITY

 

MACHINE

MODE WHEN

CHECKED

 

DISABLE WHEN EXTERNAL AG

IS IN STANDBY MODE

 

AG Hardware Error

AG module Hardware

Error.

 

Medium Runtime Yes
O2 Sensor Error Paramagnetic O2 sensor error Medium Runtime Yes

External AG Self

Test Error

 

Module fault or

communication failure

between the module and

anesthesia system

 

Low Runtime Yes

AG Hardware

Malfunction

 

AG module hardware

malfunction. The AG

module enters Standby

and measurement stops

 

High Runtime Yes
AG Init Error

The AG module was

installed improperly or

malfunctioned.

 

High Runtime Yes
AG No Watertrap

The AG module

watertrap was installed

improperly or not

installed.

 

Low Runtime Yes

AG Watertrap Type

Wrong

 

When the patient type is

infant, but the watertrap

type is adult/pediatric,

this alarm will be

triggered

 

Low Runtime Yes

AG Change

Watertrap

 

When the actual flow is

less than 75 % of the set

flow, the alarm indicates

that the watertrap is

gradually occluded and it

is necessary to replace

the water trap

 

Medium Runtime Yes
AG Comm Stop

AG module malfunction

or communication

failure.

 

High Runtime No

AG Airway

Occluded

 

Pump rate is lower than

20 ml/min for 1 second.

 

High Runtime Yes
AG Zero Failed

Gas measurements may

have bad accuracy

during zeroing

 

Low Runtime Yes

Mixed Agent and

MAC < 3

 

MAC < 3 Low Runtime Yes
Mixed Agent and MAC ≥ 3 MAC ≥ 3 Medium Runtime Yes
Mixed Agent

Two anesthetic agents

are detected but MAC is

an invalid value.

 

Medium Runtime Yes

External AG

Module

Disconnected

 

The external AG module

has been disconnected

and no O2 sensor is connected

 

High Runtime Yes

Incompatible AG

Software Version

 

The AG Version Limit is

On, and the AG module is

loaded while the AG

software version is lower

than 1.7.3.0

 

High Runtime No

CO2Over Range

N 2 over Range Hal Over Range

Enf Over Range

Iso Over Range

Sev Over Range

Des Over Range

O2 Over Range

Rate Over Range

 

The monitoring value

exceeds the measurable range

Low Runtime Yes

 

Mindray Anesthetic Gas (AG) Module P/N 115-043908-00 For Hospital

Similar Products

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.