Product Description

M8048A FMS-8 PHiliphhilip MP60 MP70 Patient Bedside Monitor Module Rack

 

 

 

 

PHiliphhilip FMS module rack

 

MAX weight:<3500g ,<7.7 lb

Weight*Height *Width:<315mm*130mm*120mm (12.4*5.1*4.7inch )

 

Flexible Module Server Troubleshooting

 

Symptoms Cause of Failure Failure Isolation and Remedy
Prompt Message “Unrecognized Measurement Module in slot s” is issued An unsupported module has been plugged into the Flexible Module Server Unplug the unsupported module
Prompt message “Measurement Module in slot n is currently ignored” is issued Too many modules of the same kind have been plugged into the Flexible Module Server Unplug module in slot n
Red Error LED stays on Unrecoverable hardware selftest error: Try connecting the MMS directly to the MSL cable. If the measured value is displayed on the screen, the fault is in the FMS
MSL cable defective If no measurements are displayed when the MMS is directly connected to the MSL cable, replace the MSL cable
Flex connector from main board to MSL defective Replace connector
CPU module defective Replace CPU module
Red Error LED flashes Hardware selftest error If system comes up, check status log. Otherwise see above
Flexible Module Server LEDs OK, Front Measurement Module not recognized (silent or INOP) Damaged measurement module or measurement module connector Replace Measurement Module
  No Front-End power because MSL voltage from the monitor is too high or too low

1. Try a new MSL cable. Replace if failure is rectified.

2. Replace mother board

  Mother board or connector on Mother Board defective If the voltage is within range, or if the connector is visibly damaged, replace the motherboard

 

 

 

 

PHiliphhilip FMS/MMS Module Lock

 

 

The following parts and hardware are included in this mounting kit

 

Item # Description Qty
1 Main Enclosure 1
2 Support Bracket 1
3 #8-32 x 1/4'' Button Head Machine Screw, Tamper-proof 4
4 5/32'' Hex Wrench, Tamper-proof 1

 

 

Mounting Notes: The bracket is designed to accommodate FMS (handle on both ends) or MMS extension (handle on one end). Installation is only shown on FMS with MMS extension.

 

 

Installing the Module Lock

 

1. After installing the required modules, slide the FMS into the main housing as shown below. Installation Note: It is not necessary to remove the previously installed mounting clips from the FMS to install the module lock.

 

 

2.Slide the support bracket over the MMS extension and into the main housing as far as it will go. Note the slot on the bracket for mounting the module lock on the second handle.

 

3.Using the 5/32'' tamper-resistant hex key provided, install four (4) #8-32 x 1/4'' tamper-resistant screws as shown below.

Installation Notes: Module locks are designed to provide space for individual modules to be disconnected without removing them from the server.

 

 

 

 

 

 