China World Technology Medical Equipment Service Group
                                                                                                           
Verified Supplier
11 Years
Since 2015
Menu

PHiliph M2735A Toco Transducer For valon F20 FM30 Fetal / Maternal Monitor

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 3-5 working days
Brand: PHiliph
Product Description

 

PHiliphhilip M2735A Toco+ Transducer for PHiliphhilip Avalon F20 FM30 Fetal/Maternal Monitor

 

 

 

► Brief Introduction

PHiliphhilip M2735A Toco+ Transducer can work for PHiliphhilip Avalon F20 FM30 Fetal/Maternal Monitor.

Several parameter frontends are combined on one board. In addition to the Toco frontend, additional supported parameters are DECG, MECG and IUP. A seven-pin ‘D-type’ socket carries all parameter related inputs and outputs. An external mode resistor, connected to one of the pins, automatically detects which mode to set when an adapter cable is plugged in (whether it is DECG, MECG, or IUP).

 

 

Environmental

 

Temperature Range Operating 0°C to 40°C (32°F to 104°F)
Storage -20°C to 60°C (-4°F to 140°F)
Humidity Range Operating <95% relative humidity @ 40°C/104°F
Storage <90% relative humidity @ 60°C/140°F
Altitude Range Operating -500 to 3000 m/-1640 to 9840 ft.
Storage -500 to 13100 m/-1640 to 43000 ft.

 

 

Transducer Cable Assembly

 

Transducer Cable (for all fetal transducers) 1
Sealing Gasket 1
Screw Shoulder M2.5 3
Screw Cover (pack of 3) 1
Transducer Belt Button 1

 

 

► Making Connections

  1. Depending on the equipment you are using, ensure the Toco+ transducer is connected to the fetal monitor.
  2. Connect a pre-gelled foam electrode to each of the two leads on the MECG adapter cable.
  3. Apply the foam electrodes to the patient, following the instructions supplied with the foam electrodes.
  4. Depending on the equipment you are using, connect the pink connector plug on the MECG adapter cable to the pink connector on the Toco+ transducer.

 

Testing the Toco+ Transducer (M2735A): DECG Mode

  1. Switch on the monitor and the recorder.
  2. Connect the patient module or Toco+ transducer to the fetal monitor.
  3. Attach the DECG adapter cable M1362B to the socket on the patient module or Toco+ transducer.
  4. Ensure that the DFHR channel display on the fetal monitor shows the DECG LEADS OFF INOP with the DECG adapter cable attached.
  5. Take a Fetal Scalp Electrode, and connect it to the DECG adapter cable.
  6. EITHER

Make a short between the spiral electrode and the reference electrode with your fingers (it is best to wet your fingers first). Use a sterile Fetal Scalp Electrode.

OR

Cut off the plastic tip of the fetal scalp electrode (containing the spiral and reference electrodes) from the end of the wires. Strip the insulation from the end of the wires, and connect them to a patient simulator.

 

 

 

 

Thank you for your time! If you are interested in PHiliphhilip M2735A Toco+ Transducer for PHiliphhilip Avalon F20 FM30 Fetal/Maternal Monitor, welcome to place your orders. Early order early shipment! 

If you have any questions, please feel free to contact us at any time, we are glad to answer them.

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.