Humidity Control High Altitude Test Chamber Altitude Low Air-Pressure Testing

Price Negotiable
Price: To Be Negotiated
MOQ: 1
Delivery Time: 30-35 business days after received the payment
Brand: Envsin
Product Description

Humidity Control High Altitude Test Chamber Altitude Low Air-Pressure Testing

 

The high altitude test chamber can be used for the heat dissipation test sample and non-heat dissipation test sample test. For the test of heat dissipation test sample, the heat dissipation power can not exceed the cooling capacity of the refrigeration system of the test chamber, because the control of temperature and humidity is a dynamic balance process, in different temperature and humidity points, the cooling capacity of the refrigeration system is also accompanied by changes.

 

Standard:
•GB/T 2423.1 Low temperature test methods •GJB 150.2 Low pressure (height) test
•GB/T 2423.2 High temperature test method •GJB 150.3 high temperature test
•GB/T2423.21 Low pressure test method •GJB 150.4 low temperature test
•IEC60068-2 Temperature/low pressure test method •GJB 150.6 Temperature - height test

 

Specifications:
Temperature Chambers: ETQ7200
Climate Chambers: ECQ7200
Test Space Volume (L) 2000
Temp. Test Parameters:
Temp. Range (℃) -70~+150
Temp. Constancy (℃) ±0.1-±0.8 (Atmospheric Pressure No Load)
Temp. Homogeneity(℃) ±0.5-±2.0 (Atmospheric Pressure No Load)
Temp. Accuracy (℃) ±2.0 (Atmospheric Pressure No Load)
Heat Up Time (min) (+ 20℃ to ~150℃)

60

Cooling Down Time (min) (+ 20℃ to~60℃)

80

Pressure Range (Kpa) Atmospheric Pressure ~1.0 Atmospheric Pressure ~0.5
Pressure Accuracy (Kpa) ±2 (Atmospheric Pressure ~40KPa); ±5%(40KPa~4KPa); ±0.1(4KPa~1KPa)
Pressure Release Time (min)

45(Atmospheric Pressure→1KPa)

Rapid Release of Pressure Function (optional)

Pressure Release Range: 75.2KPa → 18.8KPa; Pressure Release Time: 15sec

Climate Test Parameters for ECQ series:
Temp. Range (℃) +10~+95
Temp.Constancy (℃) ±0.1~±0.8 (Atmospheric Pressure No Load)
Temp. Homogeneity (℃) ±0.1~±2.0 (Atmospheric Pressure No Load)
Humid. Range (%RH) (10) 20-98 (Atmospheric Pressure No Load)
Humid. Constancy (%RH) ±1-±3 (Atmospheric Pressure No Load)
Test Space Dimension (mm)

1400W*1200D*1200H

External Dimension (mm) 1810W*3710D*2310H
Power AC380V±10%, 50HZ, 3/N/PE
Cooling Methods Water-Cooled

Options Accessories

Standard Version

• Iindependent sensors for specimen protection (NE60519-2.1993)

•1 Set testing electrode

•Nitrogen gas auxiliary device

•Electronic temperature and humidity sensors

•Spare parts package

 

•E-management and cyber-software

 

•Video monitor system

 

•APP for mobility management

 

•Rapid release of pressure device

 

•Rapid recovered pressure device

 

 

Structure:

Integral, (high altitude test chamber door direction is in the front) for the studio, refrigeration unit for the lower, after the electric control cabinet and electric control device, the main control interface is located in front of the high altitude test chamber door, the expansion control panel is located in the right back of the high altitude test chamber. Wide open door, back and side of the studio for circulating air duct system and air conditioning system: air centrifugal fan, air diversion unit, low pressure heat radiation unit, heater, humidifier, refrigeration heat exchanger, temperature/low pressure sensing unit.

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Envsin Instrument Equipment Co., Ltd.
Location 139, Yinxing Technology Building, No. 1235, Guangguang Road, Xinlan Community, Guanlan Street, Longhua District, Shenzhen
Contact Person Jack He

Request A Quote

Please check your email address.
Your message must be at least 20 characters.