ESR Series Solar Test Chambers with Integrated Solar Testing Systems
Laboratory Solar Simulation Chamber Accelerated Weathering Resistance Testing
The solar simulation chamber can be used to evaluate the ability of equipment directly exposed to solar radiation in hot season to withstand the thermal effect or photochemical effect produced by solar radiation.
The solar simulation chamber can test the heat dissipation test samples and non-heat dissipation test samples. For the test of heat dissipation test sample, the heat dissipation power can not exceed the cooling capacity of the refrigeration system of the test chamber, because the control of temperature and humidity is a dynamic balance process, in different temperature and humidity points, the cooling capacity of the refrigeration system is also accompanied by changes.
- The volume of the solar simulation test chamber shall be at least 10 times the volume of the test sample
- The light source in the solar simulation test chambe is at least 0.76m away from any surfaceof the test sample
- Test chamber sample surface and test room any of the inner wall surface distance is not less than 0.3m, the test sample should be placed to ensure that the airflow is not blocked and shielded each other.
|
Temperature test parameters :
|
Temperature and humidity climate test parameters :
|
- GJB150.7-86 solar radiation test
- Gjb150a.7-2009 Solar radiation test
- IEC60068-2-5 test Sa: surface solar radiation simulation
- GB2423.24 test Sa: Simulate the solar radiation on the ground
- DIN75220 automotive components in the sun simulation device changes
- Mil-STD 810G environmental engineering related matters and laboratory testing
- Chamber size :1100*950*1500mm(WIDTH * Depth * height)
- Overall body size: 1650*2700*2000mm(WIDTH * Depth * height)
- Inner chamber material: SUS304 high-grade stainless steel plate
- Material of outer chamber: cold rolled steel plate + double-sided electrostatic spraying process
- Insulation material: rigid polyurethane foam insulation layer
- Light wavelength: full band spectrum, simulating sunlight from 280nm to 3000nm
- Irradiation intensity :(0-1200) W/ m2 can be adjusted, and the test process can be recorded continuously
- Irradiation uniformity: 4㎡ within the uniformity of ± 10%