Overview

Basic Info.

Model NO.
WJ-1500A
Customized
Customized
Condition
New
Product Name
Chemical Fume Hood
Worktop
20+6 mm Ceramic
Liner Material
Ceramic Fiber Board
Input Power
380V/50A
Face Velocity
0.4-0.6 M/S
Application
Environment/Institute/Biology Lab/Chemical Lab
Transport Package
Standard Export Wooden Case Packing
Specification
1500*1205*2400 MM
Trademark
Ample
Origin
Chengdu, China
HS Code
8414809090
Production Capacity
200 Set/Month

Product Description

Product Description

Fume hoods are both a protective and control device fitted in laboratories where chemicals are used. They are a conventional fume extraction unit designed so that airborne contaminants are contained in the enclosure prevented from entering the room or laboratory by means of a protective air barrier between the user and the materials placed within the equipment.

Performance testing is essential to confirm that the fume hood is operating to a level to provide the desired arrest, containment and removal of the fume.

The tests are based on ANSI/ASHRAE Standard 110, Method of Testing Performance of Laboratory Fume Hoods. When the hood is already operational in the laboratory, it is recommended to test in two operating conditions.

For the "as installed" test, the laboratory hood is installed at the customer's location and tested empty, but with the ventilation system in the installation balanced and the hood in its final location. For the "as used" test, the assessment is conducted after the hood has been installed and used by the chemist. The equipment and tools remain in the hood and other activities in the laboratory continue.

Product Parameters
 
                Model Specification  WJ-1500AWJ-1500BWJ-1800AWJ-1800B
External dimensions of equipment(mm)1500(W)*1205 (D) *2400 (H)1800(W)*1205 (D) *2400 (H)
Dimension of works pace (mm)1260(W1)*780(D1) *1100 (H1)1560(W1)*780(D1) *1100 (H1)
Panel material20+6mm thick butterfly ceramics
Material of internal lining board5mm thick ceramic fiber board
Diversion structureLower air return
Control systemButton control panel (LCD panel)
PH value controlThe medium is alkaline water solution; manual monitoring, and manual control through acid pump and alkali pump.
Input powerThree-phase five-wire 380V/50A
Current for air fanNot over 2.8A(380V or 220V can be directly connected)
Maximum load of socket12 KW(total of 4 sockets)
Water tap1 set (remote control valve + water nozzle)No1 set (remote control valve + water nozzle)No
Water discharge wayMagnetic chemical pump strong discharge
Using environmentFor non-explosion indoor use, within 0-40 degrees Celsius.
Applicable fieldsInorganic chemistry experiment; Food, medicine, electronics, environment, metallurgy, mining, etc.
Ways of PurificationSpray sodium hydroxide solution, no less than 8 cubic meters/hourSpray sodium hydroxide solution.no less than 12 cubic meters/ hour
Ways of surface air speed controlManual control (through the electric air valve to adjust the exhaust air volume or adjust the height of the moving door)
Average surface air speed0.6-0.8 m/s Exhaust air volume: 1420-1890m3/h (when door height h =500mm)0.6-0.8 m/s Exhaust air volume: 1760-2340m3/h (when door height h =500mm)
Speed deviation of surface airNot higher than 10%
The average intensity of illuminationNot less than 700 Lux; Standard white and uv-free yellow LED lamps; The illumination is adjustable.
NoiseWithin 55 decibels
Flow displayWhite smoke can pass through the exhaust outlet, no overflow.
Safety inspectionNo spikes, edges; Charged body and the exposed metal resistance is greater than 2 mQ; Under 1500V voltage, no breakdown or flashover occurred for 1min test.
Resistance of exhaust cabinetLess than 160 pa
Power consumptionLess than 1.0kw/h (excluding power consumption of fans and external instruments)Less than 1.2kw/h (excluding power consumption of fans and external instruments)
Water consumptionLess than 3.2L/ hLess than 4.0L/ h
Performance of wind compensationWith a unique wind compensation structure, the volume of the wind will not cause turbulence in exhaust cabinet and will not directly blow to the staff (need to connect to the air compensation system of the laboratory)
Air volume regulating valve315mm diameter flanged type anti-corrosion electric air flow regulating valve (electric contact actuator)
 
More About the Fume Hood

There are a few types of filters that could be required in a fume hood and selecting the correct type of filters for your application is imperative to protecting yourself and your colleagues from any hazardous substances in the air. The most appropriate type of filter is dependent on what types of chemicals you will be using in your daily handlings.

Activated carbon filters are created by steam heating charcoal and graphite at high temperatures (up to 980°C). This separates the layers of crystalline carbon and creates a large internal volume which allows for adsorption of chemical particles to occur. The activated carbon is then impregnated by pre-adsorption of a chemical that will react with the pollutant that the filter is intended to trap. Four different types of activated carbon filters are commonly provided. These are filters that can trap organic, ammonia, formaldehyde and acid vapours.

Hazardous particulates are filtered and trapped in different types of filters than chemical vapours. Usually this would involve a HEPA (high efficiency particulate absorbing) or ULPA (ultra low particulate air) filter. Both filters use layers of dense fibres to create a fine mesh filter that removes contaminants as air is forced through them. HEPA filters have a 99.97% filtration efficiency for particles down to 0.3 μm and the ULPA filters have a 99.97% filtration efficiency for particles down to 0.12 μm

Detailed Photos
          
Chemical Heat Resistant Explosion Proof Acid & Alkali Resistant Fume Hood for Laboratory
Chemical Heat Resistant Explosion Proof Acid & Alkali Resistant Fume Hood for Laboratory
 
Chemical Heat Resistant Explosion Proof Acid & Alkali Resistant Fume Hood for Laboratory
Chemical Heat Resistant Explosion Proof Acid & Alkali Resistant Fume Hood for Laboratory
 
Fume Hood Maintenance

Hoods should be evaluated by the user before each use to ensure adequate face velocities and the absence of excessive turbulence.

• In case of exhaust system failure while using a hood, shut off all services and accessories and lower the sash completely. Leave the area immediately.

 Fume  hoods should  be certified, at  least annually, to ensure they are operating safely. Typical tests include face velocity measurements, smoke tests and tracer gas containment. Tracer gas containment tests are especially crucial, as studies  have shown that face velocity is not a good predictor of fume hood leakage.

• Laboratory fume hoods are one of the most important used and abused hazard control devices. We should understand that the combined use of safety glasses, protective gloves, laboratory smocks, good safety practices, and laboratory fume hoods are very important elements in protecting us from a potentially hazardous exposure.

 Laboratory fume hoods only protect users when they are used properly and are working correctly. A fume hood is designed to protect the user and room occupants from  exposure to vapors,  aerosols, toxic  materials,  odorous,  and  other  harmful substances. A secondary purpose is to serve as a protective shield when working with potentially explosive or highly reactive materials. This is accomplished by lowering the hood sash.
FAQ

6 Questions to Ask When Buying a Fume Hood:

-Which chemicals will you use within the hood?

-Is a ducted or ductless hood best suited to your needs and available space?

-Where will you place the fume hood in the lab? Consider workflows, access to external exhaust systems, and competing air patterns.

-What size fume hood will best suit your needs? Be sure to consider what (if any) equipment will be enclosed in the hood.

-Are any service fixtures or accessories such as airflow monitors, electrical outlets, water, or gas fixtures required?

-Are base cabinets for acid, solvent, or non-chemical storage required?

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Chengdu Ample Import And Export Co., Ltd.
Location T2-715, Landmark Plaza, High-Tech Zone, Chengdu, China
Contact Person Mike Ma

Request A Quote

Please check your email address.
Your message must be at least 20 characters.