Laboratory Borate Affinity POCT Hba1c Analyser OEM ODM
HbA1c Test Analyzer Hemoglobin Detection Analysis LD-100 Poct HbA1c Fully Automatic Testing Analyzer
Intended Use
HbA1c Analyzer LD-100 is One Step HbA1c Testing Analyzer, it is intended for the content etermination of hemoglobin A1c (HbA1c) in human whole blood, It's a fully automatic HbA1c testing analyzer, one step operate process without extra instrument. LD-100 is very small and light than other analyzer in same type so it can be suitable for the detection of HbA1c in medical units at all levels.
Specifications
| Name | LD-100 (POCT) HbA1c Testing Analyzer |
| Sample Type | Whole blood |
| Sample Volume | 4ul |
| Testing Time | Within 3 minutes |
| Data storage | 7000 |
| Power connection | Detachable power supply cord |
| System Compose | Sample cup mixing, Colorimetric detection And System control, Result display |
| Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
| Device category: | Fixed |
| Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
| Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
| Voltage | 12VDC-3.33A |
| Analytical wave-length | 510nm and 580nm. |
| Certification | NGSP, IFCC |
PRINCIPLE
- Uses the (borate affinity) fluorescent immunoturbidimetry
- Push a mixture containing a fluorescent borate conjugate into an integrated test cup and mix to dissolve sufficiently.
- The blood samples in the sample collector are then pushed into the buffer and mixed to release the hemoglobin from the red blood cells.
- The total concentration of hemoglobin depends on the initial reduced fluorescence signal.
- Fluorescent borate conjugates bind to glycosylated hemoglobin, and glycosylated hemoglobin is determined by detecting the decrease of fluorescent active ingredients.
- The percentage of glycosylated hemoglobin in total hemoglobin can be determined.
Feature
- One-Step Testing,Fully Automatic System
- Category of the test product: Laboratory
- Fast test speed and High Accuracy within 3 min
- Working environment: Standard
- Cv<3%, Excellent Correlation With HPLC
- Input power: <60VA
- Color touch screen just need to clik the screen and easy to read
Reliable
| IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
| Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
| CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge
Performance Parameters
- Accuracy: Tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
- Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
Analysis Interface
Click the Analysis button on the main interface to enter the sample detection program. For details, see 5.3 Sample Detection.
Certificated