HPLC Automated Hematology Analyzer For Glycosylated Hemoglobin Diaglosis
Price:
USD7,000.00-12,000.00/Unit
MOQ:
1 Unit
Delivery Time:
15-20 working days
Brand:
LABNOVATION
Product Description
HPLC Automated Hematology Analyzer For Glycosylated Hemoglobin Diaglosis
Intended use
The Fully Automated HbA1c Analyzer is used in conjunction with suitable reagents for the quantitative detection of glycosylated hemoglobin (HbA1c) in human blood.
Principle
Using high performance liquid chromatography (HPLC), the stationary phase used is a weakly acidiccation exchanger. Glycosylation of hemoglobin will result in the loss of cations on the surface of hemoglobin without being charged, and non glycosylated hemoglobin has a positive charge. According to the different charged properties of hemoglobin, eluents with different pH values and ionic strengths are used sequentially, and several hemoglobin components including HbA1c are separated by high performance liquid chromatography. Each component is detected by a single wavelength colorimeter, and the detected data is processed by a computer to calculate the percentage content of HbA1c in the total hemoglobin.
Specification
| Instrument | Manufacturer | Lanovation |
| Manufacture place | China | |
| Model | LD-560 | |
| Dection information | Principle | HPLC |
| Detection method | Dual wavelengh colorimetry | |
| Wavelength | 420nm | |
| Test time | 90 Seconds | |
| Test Speed throughtput | 40 samples per hour | |
| Sample information | Sample types | Whole blood or Pre-diluent sample |
| Test blood volume | 5μL in whole blood mode 150μL in pre-diluted mode |
|
| Sample capacity | 20 samples | |
| Sample loading method | Continuous automatic sample loading | |
| Mixer | No, but probe mixing samples | |
| Reagent and other consumble | Column test quantity | 3000 |
| Reagent pack | 600T(Reagent)/ 3000T(Column) separate | |
| Eluent | 2 | |
| Operation and character information | Operation screen | 10 inch Full color LCD display |
| Data storage | 5000 | |
| Built-in Thermal Printer | Yes | |
| Performance information | CV | <2% |
| Test range | 3.0%~18.0% | |
| Package Information |
Dimensions | 378(L) ×380(W)×510(H)mm |
| Weight | 28kg |
Product Composition
Main Parts
The analyzer consists of an auto sample injection module, a liquid chromatography separation module, a colorimetric detection and control module, a barcode scanning module, and a display printing module.
Reagents and Consumables
The analyzer is recommended to be used in conjunction with the Glycosylated Hemoglobin Test Reagent Kit (high performance liquid chromatography (HPLC method)) produced by our company, which includes the following components:
| Eluent A/Eluent B/Eluent C(Variant Mode) | Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin. |
| Hemolysis Reagent | Used to dissolve red blood cells and release hemoglobin. |
| Chromatography Column | Used for separation and elution of glycosylated hemoglobin and non glycated. |
| Hemoglobin Filter | Used to initially filter the sample after hemolysis. |
| Paper Rolls | Prints a report of the test results. |

Exhibition