Low Noise VPSA Oxygen Generator with Long Maintenance Cycle and Modular Design for Industrial Use
- Low power consumption (0.32-0.38kWh/Nm³)
- Fast start-up capability
- Simple maintenance requirements
- Modular design for flexible production capacity
| Model No. | Flow Rate (Nm³/h) | Purity |
|---|---|---|
| VPSA-5 | 5 | 93±2% |
| VPSA-10 | 10 | 93±2% |
| VPSA-20 | 20 | 93±2% |
| VPSA-30 | 30 | 93±2% |
| VPSA-40 | 40 | 93±2% |
| VPSA-50 | 50 | 93±2% |
| VPSA-60 | 60 | 93±2% |
| VPSA-70 | 70 | 93±2% |
| VPSA-80 | 80 | 93±2% |
| VPSA-100 | 100 | 93±2% |
| VPSA-120 | 120 | 93±2% |
| VPSA-150 | 150 | 93±2% |
| VPSA-180 | 180 | 93±2% |
| VPSA-200 | 200 | 93±2% |
Before installing a VPSA Oxygen Generator, ensure the site meets these critical requirements:
- Structural Requirements: Flat, solid surface with floor bearing capacity of 5kN/m² (small units) to 15kN/m² (large systems). Foundation must use C30 concrete with steel plates (≥12mm thickness) for equipment fixation.
- Environmental Conditions: Well-ventilated area (≥3 air changes/hour) with relative humidity ≤85%. Avoid dust sources or install appropriate filtration.
- Space Requirements: Minimum 1.5m clearance from walls/obstacles for maintenance access.
The installation involves three key stages:
- Equipment Positioning: Level adjustment (≤0.5mm/m deviation) and secure anchoring of blower/vacuum pump.
- Pipeline Connection: Use corrosion-resistant stainless steel pipes (304/316L) with 1‰-3‰ slope. PTFE gaskets for airtight seals followed by pressure testing (1.2× working pressure for 30 minutes).
- Electrical Installation: 380V/50Hz three-phase supply with dedicated circuit (≤30mA leakage protection). Grounding resistance ≤4Ω. PLC system integration via Modbus protocol if required.
Daily: Monitor pressure (0.1-0.3MPa) and flow (±5% of rated value). Clean air pre-filter.
Weekly: Inspect seals for leaks (soapy water test). Check blower/vacuum pump oil levels.
Monthly: Calibrate oxygen purity sensor. Test pressure relief valve functionality.
Troubleshooting: For purity below 90%, check filter and molecular sieve condition. Excessive vibration may indicate loose bolts or bearing wear. Molecular sieve replacement typically required every 5-8 years depending on usage intensity.
We also manufacture PSA Oxygen Generators, storage tanks, and heat exchangers. For inquiries about VPSA Oxygen Systems or other products, please contact us at info@gneeheatex.com.