Guangzhou BioKey Healthy Technology Co.Ltd
                                                                                                           
Verified Supplier
23 Years
Since 2003
Menu

Lyophilized 96 Tests/Kit Real Time STD STI Test Kit for Group B Streptococcus GBS

Price Negotiable
Price: USD
MOQ: We can produce liquid and lyophilized kits
Delivery Time: Depending on order quantity
Brand: Biokey
Product Description
Group B streptococcus(GBS) real-time PCR detection kit (lyophilized)
--96 tests/kit

 
Intended use:
 

This kit, adopting Real-Time PCR-Fluorescence technology, is used for detection of GBS. It can be used for the laboratory diagnosis and monitoring of GBS infection. The test results are only for clinical reference, not for case confirmation or exclusion.

 

Principle:
 

The specific primers and specific fluorescent probes are designed for the nucleic acid conserved region of GBS. The PCR reaction solution and real-time fluorescent quantitative PCR detection technology is applied on the fluorescence quantitative PCR instrument to realize the rapid detection of GBS through the change of fluorescence signal. The detection system contains a human internal control (IC) gene, which is used to monitor whether the specimens and the process of nucleic acid extraction is qualified.


 
Main components:

 

No. Components BIK-QL-H0021

1

PCR Enzyme  Mix(Lyophilized)

96 tests/bottle

2 GBS Primer-Probe Mix 100 μL/vial
3 Enzyme Mix Buffer (5×) 400 μL/vial
4 Positive Control 100 μL/tube
5 Negative Control 100 μL/tube
6 RNase-free water 1mL/tube
7 Paraffin oil 1.5mL/tube*2


Note: Do not mix the components from different batches for detection. The Positive Control of GBS and internal control were constructed artificially, and they were not infectious.  

 

GBS product lyophilization pacakage:

 

 

 

Specimen requirements:

 

1. Specimen type

Vaginal, rectal, or combined vaginal and rectal swabs of pregnant women.

2. Specimen Preservation

The specimen can be stored for 3 months at -20±5℃ and for a long time below -70℃

 

PCR Amplification condition:

 
PCR amplification according to the following procedures:
 

Step Cycle number Temperature Time
1 1 95℃ 3min
2 40 95℃ 10s
60℃ 30s collect fluorescent

 

Instrument detection channel selection: FAM and HEX/VIC.
 
Quality control:
 

  Ct value
Negative Control No Ct
Positive Control ≤30
IC of samples ≤38


Note: Negative Control, Positive Control and IC of samples must be performed correctly, otherwise the sample results is invalid.
 

Compatible instruments:

 

The diagnostic kit is applicable to PCR instrument with multi-color fluorescence channel FAM and HEX/VIC, including Quant Studio5 (ABI), Micgene 244/244 IVD (BIO-KEY).

 

 Operation details please check from IFU!

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Guangzhou BioKey Healthy Technology Co.Ltd
Location Add:No.2, Ruitai Rd, Guangzhou Hi-Tech Industrial Development Zone, 510670, CHINA
Contact Person Lisa

Request A Quote

Please check your email address.
Your message must be at least 20 characters.