Product Description

Four Insect-borne Infectious Diseases Pathogens(Malaria) Multiple Fluorescence PCR probe kit

 

Intended use:
 

This product is used for in vitro qualitative detection of Malaria/Borrelia burgdorferi/Zika/Dengue virus nucleic acid in human serum samples. The test results are for clinical reference only, it cannot be used alone as the basis for the diagnosis or exclusion of cases.

 

Test Principle:


The kit adopt TaqMan fluorescent probe method. For the highly conserved and specific region of Malaria/Borrelia burgdorferi /Zika/Dengue, the primer probes, together with RT-PCR reaction solution for fluorescent detection are designed and labeled with different fluorescent groups. On the fluorescence quantitative PCR instrument, the real-time fluorescence quantitative PCR detection technology is used to realize the rapid detection of Malaria/Borrelia burgdorferi/Zika /Dengue through the change of fluorescent signal. The detection system contains positive internal control, used to monitor whether the nucleic acid extraction process is normal.


Product components:

 

1 BIK-QL-H009:

No. Components Amount
1 PCR Enzyme Mix (Lyophilized) 96 tests/bottle
2 Primer-Probe Mix 1( Malaria ) 50 μL/vial
3

Primer-Probe Mix 2(Borrelia

burgdorferi / Zika / Dengue)

50 μL/vial
4 Enzyme Mix Buffer (5×) 400 μL/vial
5 Positive Control 100 μL/tube
6 Negative Control 100 μL/tube
7 RNase-free water 1mL/tube
8 Paraffin oil 1.5mL/tube*2

 

2 BIK-QL-H009S:

 

No. Components Amount
1 1 (Malaria Mix (Lyophilized)) 24 tests/kit
2 2 (Borrelia burgdorferi/ Zika / Dengue Mix (Lyophilized)) 24 tests/kit
3 Positive Control 100 μL/tube
4 Negative Control 100 μL/tube
5 RNase-free water 1mL/tube
6 Paraffin oil 1.5mL/tube

 

Note: Do not mix the components from different batches for detection. The Positive Control of Malaria/Borrelia burgdorferi/Zika/Dengue and internal control were constructed artificially, and they were not infectious.

 

Product pictures:

 

 

 

 


Compatible instruments:

 

The diagnostic kit is applicable to PCR instrument with multi-color fluorescence channel FAM , VIC/HEX, ROX and CY5 including Quant Studio5 (ABI), Micgene 244/244 IVD(BIO-KEY).

 

Reaction Mix Preparation:

 

Ⅰ. Malaria reaction mix:

 

1×volume required
Resuspended master mix 14 μL
Primer-Probe Mix 1 1 μL
Total volume 15 μL

 

 

PCR amplification condition:

 

Step Cycle Temperature Time
1 1 50℃ 10 min
2 1 94℃ 3 min

 

3

 

40

94℃ 10 s
60℃

30 s(Collection of fluorescence)

 

Interpretation of test results:

 

(1)Malaria reaction:

 

 

FAM

(Malaria)

CY5

(IC)

Interpretation test results
1 - - Result invalid ,test again
2 + +/- Malaria(+)
3 - + Negative(-)

 

NOTE: ‘+ ’ is positive, ‘- ’ is negative.


 Operation details please check from IFU

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Guangzhou BioKey Healthy Technology Co.Ltd
Location Add:No.2, Ruitai Rd, Guangzhou Hi-Tech Industrial Development Zone, 510670, CHINA
Contact Person Lisa

Request A Quote

Please check your email address.
Your message must be at least 20 characters.