Furation metabolite colloidal gold rapid test card tissue Aquatic products

 

 

Overview:
 
Furantoin (AHD) rapid detection card is convenient, fast and sensitive, suitable for on-site mass sample detection. The test card contains antigen in the test area (T) and anti-ⅱ in the control area (C) of nitrocellulose membrane fixed in advance, and gold label antibody fixed on the binding pad. According to T area, C area color contrast to determine the result. A purplish red band appeared in zone C regardless of the presence or absence of furantoin metabolites.
 
Detection limit:

 

Pork, liver, chicken, duck, beef, lamb, chicken skin, bullfrog, fish, shrimp meat: 0.5 PPB (µg/kg)

 

Product component:
 

Item No. Name
1 Furazolidone metabolite rapid test card
2 n-hexane
3 Reagent A
4 Reagent B
5 Reagent C
6 Reconstitution fluid
7 User manual

 
Preservation and stability:

 

Store in dry place at 20±5℃, do not freeze, avoid direct sunlight.

This test card is valid for 12 months, and the production date is shown in the packaging box.

 
Materials required but not provided:

 

Item No. Name
1 Homogenizer
2 Vortex Oscillator
3 Electronic Balance
4 Centrifuge
5 Nitrogen (air) blower
6 Micropipette
7 water bath pot
8 Deionized water

 

Detection Procedure:

 

1. Restore the test card and sample solution to room temperature before use;

2 .Take out the detection card and place it flat on the desktop;

3.  Absorb the lower layer solution of the sample to be tested 80 μL, add vertically to the sample adding well;

4 . 5 min after adding samples, observe the color area and determine the result.

 

Result Judgment:

 

Negative: C line color, T line darker than C line color or consistent,The test results were negative.

Positive: C line is colored, T line is lighter than C line or T line is light green, and the test result is positive.

Invalid: Line C does not display color, regardless of whether line T displays color, the detection card is judged to be invalid.

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Guangzhou BioKey Healthy Technology Co.Ltd
Location Add:No.2, Ruitai Rd, Guangzhou Hi-Tech Industrial Development Zone, 510670, CHINA
Contact Person Lisa

Request A Quote

Please check your email address.
Your message must be at least 20 characters.