Furazolidone metabolite colloidal gold rapid test card tissue, Aquatic products
 

 

Overview:
 

Furazolidone (AOZ) rapid detection card is convenient, fast and sensitive, suitable for on-site mass sample detection. The test card contains antigen in the test area (T) and anti-ⅱ in the control area (C) of nitrocellulose membrane fixed in advance, and gold label antibody fixed on the binding pad. According to T area, C area color contrast to determine the result. Whether or not furazolidone metabolites were present in the samples, a purplish red band appeared in the C region.

 
Detection limit:

 

Pork, liver, chicken, duck, beef, lamb, chicken skin, bullfrog, fish, shrimp meat: 0.5 PPB (µg/kg)

 
Product component:
 

Item No.Name
1Furazolidone metabolite rapid test card
2n-hexane
3Reagent A
4Reagent B
5Reagent C
6Reconstitution fluid
7User manual

 
Preservation and stability: 
 

Store in dry place at 20±5℃, do not freeze, avoid direct sunlight.

This test card is valid for 12 months. The production date is shown in the package
 
Materials required but not provided:
 

Item No.Name
1Homogenizer
2Vortex Oscillator
3Electronic Balance
4Centrifuge
5Nitrogen (air) blower
6Micropipette
7water bath pot
8Deionized water


Detection Procedure:

 
1. Restore the test card and sample solution to room temperature before use;
2. Take out the detection card and place it flat on the desktop;
3. Absorb the lower layer solution of the sample to be tested 80 μL, add vertically to the sample adding well;
4. 5 min after adding samples, observe the color area and determine the result.
 
Result Judgment:
 

Negative: C line color, T line darker than C line color or consistent,The test results were negative.
Positive: C line is colored, T line is lighter than C line or T line is light green, and the test result is positive.
Invalid: Line C does not display color, regardless of whether line T displays color, the detection card is judged to be invalid.
 
 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Guangzhou BioKey Healthy Technology Co.Ltd
Location Add:No.2, Ruitai Rd, Guangzhou Hi-Tech Industrial Development Zone, 510670, CHINA
Contact Person Lisa

Request A Quote

Please check your email address.
Your message must be at least 20 characters.