Technical Parameters of SFJ-231 Helium Leak Detector:

Product Name Helium Leak Detector SFJ-231
Diamension 645*678*965mm
Operating temperature 0~40°C
Detectable Quality 2, 3, 4 (H2, He3, He4)
User Interface 7inch Color Touch Screen
Detection Method Helium Leak Detection
Mini Leak Detection Rate(Pa.m3/s)/Vaccum Mode 5*10-13 Pa·m3/s
Communication Interface RS232/485, USB*2
Max Allowable Leak Detection Pressure (Pa) 1500
Ion Source 2pcs, Iridium Coated Yttrium Oxide, Automatic Switching
Mini Leak Detection Rate(Pa.m3/s)/Sniffer Mode 5*10-9 Pa·m3/s
Start Time 2min
Leak Detection Port DN25KF
Power AC220V, 50Hz/60Hz
Language Chinese/English

Helium Leak Detection for Power Plant Steam Systems (Using Wayeal SFJ-231)
Sealing integrity in power plant steam systems (e.g., turbines, steam pipelines, valves, and heat exchangers) directly impacts unit efficiency and operational safety. Even minor leaks can cause steam loss, reduced thermal efficiency, corrosion, or unplanned shutdowns. The Wayeal SFJ-231 helium mass spectrometer leak detector delivers precise, reliable detection for these critical applications.

Leak Detection Method (Vacuum Mode)
1. System Preparation

  • Isolate the steam system and evacuate it to ≤1500 Pa (maximum allowable detection pressure).
  • Connect the SFJ-231’s DN25KF leak detection port to the test point.

2. Helium Spraying

  • Use a helium spray gun (compatible with H2/He3/He4) to slowly spray potential leak points (welds, flanges, etc.).
  • The dual ion sources (iridium-coated yttrium oxide) auto-switch for real-time monitoring.

3. Leak Identification

  • Monitor real-time graphs/bar charts on the 7-inch color LCD touchscreen:
  • Audible/visual alarms trigger if leak rate exceeds 5.0×10-6 Pa·m³/s (sniffer mode threshold).
  • Vacuum mode detects ultra-fine leaks (as low as 5.0×10⁻13 Pa·m³/s).

4. Data Recording

  • Export results via RS232/485 or USB to MES systems for compliance reporting.
  • Recheck & Maintenance

5. Recheck & Maintenance

  • Retest after repairs to confirm leak rates meet standards.
  • Clean ion sources post-use to prolong lifespan.

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Anhui Wanyi Science and Technology Co., Ltd.
Location No. 8, Wenqu Road, High-tech Zone, Hefei, Anhui
Contact Person Annie

Request A Quote

Please check your email address.
Your message must be at least 20 characters.