EMIRS200-AT01T-BR090 Low-Power Portable Infrared Light Source For Portable Gas Analyzers
EMIRS200-AT01T-BR090 Low-Power Portable Infrared Light Source For Portable Gas Analyzers
Product Overview:
The EMIRS200-AT01T-BR090 is a compact, high-efficiency infrared (IR) light source designed for portable analytical instruments, gas sensors, and spectroscopic applications. Leveraging MEMS-based blackbody radiation technology, it delivers stable, broadband IR emission across 2–14 μm with low power consumption. The integrated TO-39 metal package with a reflective surface (BR090) optimizes radiation distribution for directional output, making it ideal for handheld devices and field deployable systems.
Key Features:
- Low-Power Operation
- Typical power consumption: ≤ 170 mW (pulsed mode) .
- Efficient thermal management minimizes heat dissipation, extending battery life in portable setups.
- Portable Design
- Compact TO-39 housing (Ø10 mm × 12 mm) with integrated reflector for directional emission.
- Lightweight (≈5 g) and rugged, suitable for mobile applications in harsh environments.
- Blackbody Radiation Performance
- Emissivity >0.95 across 2–14 μm, mimicking ideal blackbody behavior .
- Spectral uniformity ensures consistent results in gas analysis (e.g., CO₂, VOCs) and material characterization.
- Fast Modulation
- Supports pulse-width modulation (PWM) up to 50 Hz with >80% modulation depth .
- Enables advanced techniques like photoacoustic spectroscopy (PAS) and non-dispersive infrared (NDIR) sensing.
- Environmental Resistance
- Operating temperature range: -25°C to +60°C .
- IP67-rated housing withstands dust, moisture, and vibration .
Applications:
- Portable Gas Analyzers
- Detection of CO₂, methane, and volatile organic compounds (VOCs) in handheld devices.
- Used in environmental monitoring, industrial safety, and medical breath analysis .
- Spectroscopic Instruments
- Enables compact FTIR spectrometers and NDIR sensors for field applications.
- Supports material analysis (e.g., plastics) and food quality control arcoptix.com.
- Automotive & HVAC
- In-vehicle emission testing and refrigerant leak detection.
- Ventilation system optimization via CO₂ sensing .
- Medical Devices
- Non-invasive blood gas monitoring and anesthesia gas analysis .
Driving Recommendations:
- Pulsed Operation: Use PWM with 10–50 Hz frequency and 5–90% duty cycle to minimize power consumption while maintaining signal integrity .
- Thermal Management: Integrate heat sinks or thermoelectric coolers (TECs) for prolonged operation in high-temperature environments.
|
Parameter |
Specification (Typ. / Max. / Min.) |
Test Conditions |
|---|---|---|
| Part Number | EMIRS200-AT01T-BR090 | - |
| Product Type | Low-Power Portable Infrared (IR) Light Source | - |
| Core Technology | MEMS-Based Blackbody Radiation | - |
| Spectral Emission Range | 2 – 14 μm | Broadband IR coverage |
| Peak Emission Wavelength | ~9.6 μm | Optimized for CO₂ absorption applications |
| Radiant Power | 33 mW | At 500°C operating temperature; measured at source output |
| Emissivity | >0.95 | Across 2 – 14 μm spectral range |
| Power Supply Voltage | 5 – 12 V DC | - |
| Power Consumption | ≤ 170 mW | Pulsed mode (10 Hz frequency, 62.5% duty cycle) |
| Forward Voltage | 2.5 V | At typical operating current |
| Operating Current | 68 mA | At 2.5 V forward voltage; 500°C operating temperature |
| Modulation Capability | - Modulation Frequency: Up to 50 Hz - Modulation Depth: >80% |
Pulse-Width Modulation (PWM) |
| Response Time (Rise/Fall) | < 10 ms | From standby to operating temperature (rise); from operating temp to standby (fall) |
| Operating Temperature Range | -25°C – +60°C | No condensation; stable emission performance within range |
| Storage Temperature Range | -40°C – +85°C | - |
| Housing Type | TO-39 Metal Package (with BR090 Reflective Surface) | - |
| Housing Dimensions | Ø10 mm × 12 mm (Diameter × Length) | Approximate; excludes lead wires |
| Weight | ≈ 5 g | Including package and internal components |
| Ingress Protection Rating | IP67 | IEC 60529 standard; dust-tight and protected against temporary submersion |
| Vibration Resistance | 10 – 500 Hz, 20 G (X/Y/Z axes, 1 h each) | IEC 60068-2-6 standard |
| Shock Resistance | 1000 G, 0.5 ms duration (X/Y/Z axes, 3 times each) | IEC 60068-2-27 standard |
Get in Touch