ShenzhenYijiajie Electronic Co., Ltd.
                                                                                                           
Verified Supplier
20 Years
Since 2006
Menu

CH2O/C-10 Formaldehyde Gas Sensor 3 Electrode Electrochemical Highly Sensitive CH2O Measurement

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 3-5Workingdays
Brand: MEMBRAPOR
Product Description

CH2O/C-10 Formaldehyde Gas Sensor 3 Electrode Electrochemical Highly Sensitive CH2O Measurement

 

Applications
• Continuous Air Quality Monitoring
• Safety and Environmental Control

 

Formaldehyde Gas Sensor CH2O/C-10
Cross Sensitivity Data
The table below does not claim to be complete. Interfering gases should not be used for
calibration. Please contact Membrapor AG for further support regarding cross sensitivities.

 

Specification Sheet

Rec. Load Resistor 10 - 33 Ω
Bias (V_Sens-V_Ref) not recommended
Relative Humidity Range 15 % to 90 % RH non-condensing
Temperature Range -40 °C to 50 °C
Pressure Range Atmospheric ± 10%
Pressure Coefficient N.D.
Humidity Effect Abrupt changes in RH causes a short term transient signal
Expected Operation Life 3 years in air
Expected Long Term Output Drift in air < 2 % signal loss per month
Storage Life 6 months in container
Rec. Storage Temperature 5°C - 20°C
Warranty Period 12 months from date of dispatch

 

 

 