Product Description
Recycle Fluke 5720A Precision Multi Function Calibrator Wideband Output
Product Overview
The Fluke 5720A Precision Multifunction Calibrator is a five-function calibrator designed to handle demanding electrical calibration workloads. This recycled unit sources direct and alternating voltage and current, plus resistance. The wideband voltage option extends overall AC bandwidth from 10 Hz to 30 MHz to cover RF voltmeters. Compatible with 5725A and 5205A/5215A amplifiers.
Fluke 5720A Precision Multifunction Calibrator front view
Key Specifications
  • Minimum Output: 0V for all ranges (except 100V for 1100V range)
  • Maximum Load: 50 mA for 2.2V through 220V ranges; 20 mA for 1100V range
  • Load Regulation: <0.2 ppm + 0.2 mV change, full load to no load
  • Line Regulation: <0.1 ppm change, ±10% of selected nominal line
  • Settling Time: 3 seconds to full accuracy; +1 second for range or polarity change
  • Overshoot: <5%
  • Common Mode Rejection: 140 dB, DC to 400 Hz
  • Remote Sensing: Available 0V to ±1100V, on 2.2V through 1100V ranges
Wideband Output Details
  • dBm Reference: 50Ω
  • Minimum Output: 300 µV (-57 dBm)
  • Frequency Uncertainty: ±0.01%
  • Frequency Resolution: 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz
  • Overload Protection: Short circuit protection with automatic recovery
Environmental Specifications
  • Operating Temperature: 0°C to 50°C
  • Calibration Temperature: 15°C to 35°C
  • Storage Temperature: -40°C to 75°C
  • Relative Humidity: <80% to 30°C, <70% to 40°C, <40% to 50°C
  • Warm-up Time: 2x time since last warmed up (max 30 minutes)
Physical Specifications
  • Dimensions: 17.8 cm H × 43.2 cm W × 63.0 cm D (7" × 17" × 24.8")
  • Weight: 27 kg (62 lbs)
  • Power Requirements: 47-63 Hz; selectable nominal voltages: 100V, 110V, 115V, 120V, 200V, 220V, 230V, 240V
  • Maximum Power: 300VA
Available Options
Option Code Description
03 Wideband Output
Y5735 5.25" Rackmount Adapter
Additional Images
Fluke 5720A side view and controls Fluke 5720A rear panel connections Fluke 5720A calibration certificate or documentation
We recycle and trade in used test equipment including Spectrum Analyzers, Network Analyzers, Signal Generators, Calibrators, Multimeters, Oscilloscopes and many other types.

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Shenzhen Meigaolan Electronic Instrument Co. Ltd
Location RA1-B2,F32 of Dongjianghaoyuan, Baomin Rd, Bao'an District, Shenzhen,China
Contact Person Dannie

Request A Quote

Please check your email address.
Your message must be at least 20 characters.