Philip M3001A MMS Module for MP Series Monitors
Price:
Negotiation, pls contact me
MOQ:
1
Delivery Time:
Within 1-2 working days after the payment was confirmed
Brand:
Philip
Product Description
M3001A Multi-Measurement Module (MMS) - Complete Technical Specification
Overview
The M3001A is a Multi-Measurement Module (MMS) designed for IntelliVue patient monitoring systems. As the core measurement acquisition unit, it processes physiological signals and transmits data to host monitors.
Core Technical Specifications
General Information
表格
| Parameter | Specification |
|---|---|
| Device Type | Multi-parameter Multi-Measurement Module (MMS) |
| Classification | Class II Medical Device |
| Power Source | Powered by host monitor via MSL connection (no independent battery) |
| Weight | 1-3 lbs (0.45-1.36 kg) |
| Dimensions | 9" × 3.75" × 2.375" (22.86 × 9.53 × 6.03 cm) |
| Operating Altitude | -500 m to 4600 m (-1600 to 15,000 ft) |
| Storage Altitude | -500 m to 15300 m (-1600 to 50,000 ft) |
Standard Measurement Parameters
Base Configuration (Included with All Options)
-
ECG/Respiration: 3-lead, 5-lead, or 12-lead ECG monitoring
- Full arrhythmia detection and analysis
- ST segment monitoring with alarms
- QT/QTc interval monitoring
- Impedance respiration (no additional sensors required)
- Pace pulse detection
-
SpO₂ (Pulse Oximetry):
- Arterial oxygen saturation (%)
- Pulse rate (bpm)
- Real-time plethysmogram waveform
- Perfusion index
-
NBP (Non-Invasive Blood Pressure):
- Oscillometric method
- Systolic, diastolic, and mean pressure values
- Adult, pediatric, and neonatal modes
- Manual, automatic, and STAT measurement modes
Available Configuration Options
SpO₂ Technology Options
| Option Code | SpO₂ Technology Type |
|---|---|
| A01 | Fast-response SpO₂ technology (compatible with standard SpO₂ sensors) |
| A02 | Standard SpO₂ technology platform |
| A03 | Advanced motion-resistant SpO₂ technology |
| A04 | Standard SpO₂ technology platform |
Enhanced Capability Options
| Option Code | Additional Capabilities |
|---|---|
| C06 | 1 Invasive Pressure channel + 1 Temperature channel |
| C12 | 12-Lead ECG capability (conventional 10-electrode) |
| C18 | 12-Lead ECG + Invasive Pressure + Temperature |
Common Combined Configurations
- A01: ECG/Resp, Fast SpO₂, NBP
- A01C06: ECG/Resp, Fast SpO₂, NBP, IBP, Temperature
- A01C12: ECG/Resp, Fast SpO₂, NBP, 12-Lead ECG
- A01C18: ECG/Resp, Fast SpO₂, NBP, 12-Lead ECG, IBP, Temperature
- A03C18: ECG/Resp, Motion-resistant SpO₂, NBP, 12-Lead ECG, IBP, Temperature
Data Storage & Portability
Trend Storage
- Capacity: Up to 8 hours of trend data
- Resolution: 1-minute intervals for all numeric parameters
- Storage Location: Stored internally within the module (not host monitor)
Data Retention While Disconnected
- Patient demographics, trends, and settings retained for 6 hours without power
- Enables seamless patient transport between monitors without data loss
- Automatic data upload when reconnected to a new host monitor
Host Monitor Compatibility
- MP Series: MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90
- MX Series: All MX-series monitors
- M-Series Legacy: M3046A (M/M4 series, Release E and later)
Connection Methods
- Direct Attachment: Cable-free docking to rear of host monitor
- Remote Placement: Via MSL (Measurement Link) cable for bedside placement
Optional MMS Extension Modules
The M3001A supports field-attachable extensions to expand measurement capability:
表格
| Extension Model | Added Capabilities |
|---|---|
| M3015A | Sidestream CO₂ capnography |
| M3016A | Mainstream CO₂ capnography |
| M3012A | Additional invasive pressure + Temperature + CO₂ |
Technical Architecture
Internal Hardware Components
-
CPU Board:
- 68360 Controller
- Memory system (Flash ROM, SRAM, DRAM)
- NBP A/D Converters
- Real-time clock (RTC)
-
Front End Board:
- ECG/Respiration Front End
- SpO₂ Front End
- Pressure/Temperature Front End
- Floating/Non-Floating Isolation circuitry
-
NBP Pneumatic Assembly:
- Pressure pump
- Solid-state pressure sensor
- Overpressure safety system
Environmental & Performance Specifications
Temperature Measurement Range
- Range: -1°C to 45°C (30°F to 113°F)
- Isolation: Patient-applied parts isolated via opto-couplers and transformer
Invasive Pressure
- Real-time pressure waveform
- Systolic, diastolic, and mean values
- Automatic transducer detection via connector coding
Maintenance Requirements
- NBP Calibration: Required every 2 years
- SpO₂ Performance Test: Every 2 years
- CO₂ Calibration: Annually (if equipped with CO₂ extension)
- Safety Testing: Every 2 years (leakage current, protective earth continuity)
Key Clinical Features
-
12-Lead ECG Support:
- Conventional 10-electrode placement (diagnostic quality per ANSI/AAMI EC11-1991)
- EASI 12-lead derivation using only 5 electrodes
-
Advanced ECG Analysis:
- Multi-lead arrhythmia detection at bedside
- Up to 12-lead ST segment analysis with alarms
- QT/QTc interval monitoring and variation tracking
-
Fast-Response SpO₂:
- Robust performance during patient movement and low perfusion states
- Compatible with standard SpO₂ sensors including forehead sensors
-
Seamless Patient Transport:
- Module retains all patient data during transport
- No re-zeroing or reconfiguration required at destination
Regulatory Compliance
- FDA 510(k) cleared (Product Code MHX, Class II)
- CE marked
- ANSI/AAMI EC11-1991 compliant for ECG
- IEC 60601-1 compliant for electrical safety
Similar Products
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.
Company
STAR 9 BIOLOGICAL TECHNOLOGY CO.,LTD.
Location
ROOM 1199 , NO 119 JINSHAZHOU ROAD JINSHA STREET BAIYUN AREA GUANGZHOU
Contact Person
Rita