1898200T M4 Straightshot Microdebrider and Handpiece for repair
1898200T M4 Straightshot Microdebrider and Handpiece for repair
► Brief Description
Fault: 1) The bearing is corroded and the seal is broken.
2) Motor stuck, inside is full of disinfection residue, motor corrosion.
3) The rotation direction of the knife head is stuck.
4) Handle paint off, internal corrosion..
Solution: Exchange the whole handpiece except cable.
►Problem & Solutoin
| Model | 1898200T |
| Fault | 1) The bearing is corroded and the seal is broken. 2) Motor stuck, inside is full of disinfection residue, motor corrosion. 3) The rotation direction of the knife head is stuck. 4) Handle paint off, internal corrosion. |
| Solution | Exchange the whole handpiece except cable. |
► Common issues with 1898200T handpiece
- Overheating
- Mechanical Jam or Abnormal Noise
- Reduced Performance or Inconsistent Operation
- Damage to the Handpiece Tip or Blade Engagement
If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue.
► About Us
CareUtech is the global provider of medical equipment and medical equipment repair service, as well as offering spare parts. We focus on repair and repair training for monitor, Flexible & Rigid Endoscope, ultrasound, camera system, dynamical system and surgical instrument.
Shipping and packing
All products will be tested and sterilized for free, perfectly packed to ensure the safety of the goods, strictly in accordance with international export rules. We send to worldwide via EMS, DHL, FEDEX or other express, based on your choice.
► What we can do
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.