Medical Device Repair Cases
Ultrasonix EC9-5/10 Endo-cavitary probe for repair
Ultrasonix EC9-5/10 Endo-cavitary probe for repair ► Brief Description Fault: Bad array element. Solution: Replace crystal and housing. ►Problem & Solutoin BrandUltrasonix ModelEC9-5/10FaultBad array elementSolutionReplace crystal and housing ► Common issues with EC9-5/10 Probe Image Dropouts or Streaks (Dead Elements)Acoustic Lens DamageCable and Strain Relief DamageHousing or Casing Damage ►FAQHow can I tell if my EC9-5/10 has a crystal failure or a cable failure?Crystal
Aloka UST-5413 linear array ultrasound probe for repair
Aloka UST-5413 linear array ultrasound probe for repair ► Brief Description Fault: Lens break. Solution: Replace lens. ►Problem & Solutoin BrandAlokaModelUST-5413FaultLens breakSolutionReplace lens ► Common issues with UST-5413 Probe Image dropouts or black linesAcoustic lens damageCable damageConnector damageWeak or noisy image ►FAQ How can we prevent the acoustic lens on the UST-5413 from peeling after a repair? Avoid Solvent Cleaners: Never use solutions containing high
GE 4D3C-L 4D Convex Transducer For Repair
GE 4D3C-L 4D Convex Transducer For Repair ► Brief Description Fault: 1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged. Solution:Replace the dome, dome tube; repair electric motor. ►Problem & Solutoin BrandGEModel4D3C-LFault1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged.SolutionReplace the dome, dome tube; repair electric motor ►Common issues with 4D3C-L 4D
PENTAX Medical EPK-i7000 Video Processor For Repair
PENTAX Medical EPK-i7000 Video Processor For Repair ► Brief Description Brand: PENTAX (PENTAX Endoscope Processor) Model: EPK-i7000 Serial Number: AC*** Fault: It cannot recognize the endoscope and reports an error. Solution: Repair the system board. Repair Time: 3-5 working days. Warranty Period: Six months. ►Problem & Solution Brand PENTAX Model EPK-i7000 Product Processor Fault It cannot recognize the endoscope and reports an error. Solution Repair the system board. ►
C5-1 convex probe for repair
C5-1 convex probe for repair ► Brief Description Fault:Lens problem, cable broken, without strain relief. Solution:Replace probe lens, replace a new cable, replace strain relief. ►Problem & Solutoin Model C5-1 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, replace a new cable, replace strain relief ► Common issues with C5-1 probe Cable and Connector Damage Acoustic Lens Damage Crystal/Element Failure Intermittent Signal Loss or Noise
L15-7io Linear probe for repair
L15-7io Linear probe for repair ► Brief Description Fault: Lens problem. Solution: Replace probe lens. ►Problem & Solutoin Model L15-7io Fault Lens problem Solution Replace probe lens ► Common issues with L15-7io probe Cable and Connector Issues Acoustic Lens Damage Crystal/Element Failure Housing/Physical Damage Oil Leak If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue.
S5-2 Cardiac probe for repair
S5-2 Cardiac probe for repair Fault: Lens problem, cable broken, without strain relief. Solution: Replace probe lens, repair cable, replace strain relief. ►Problem & Solutoin Model S5-2 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, repair cable, replace strain relief ► Common issues with S5-2 Cardiac probe Poor or blurry images Constant noise or static Color streaking or intermittent artifacts Cracked or dried-out lens Cable or strain
L12-3 Ultrasuound probe for repair
L12-3 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Model L12-3 Fault Lens problem Solution Replace lens ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket Damage Connector Damage Strain Relief Separation Membrane/Lens Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue.
L12-3 linear Probe for repair
L12-3 linear Probe for repair ► Brief Description Fault: Array element aging, lens bubble. Solution: Replace Crystal. ►Problem & Solutoin Model L12-3 Fault Array element aging, lens bubble Solution Replace Crystal ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket Damage Connector Damage Strain Relief Separation Membrane/Lens Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will
S5-1 sector array ultrasound probe for repair
S5-1 sector array ultrasound probe for repair ► Brief Description Fault: Array element aging, lens bubble. Solution: Replace Crystal. ►Problem & Solutoin Model S5-1 Fault Array element aging, lens bubble Solution Replace Crystal ► Common issues with S5-1 Probe Image Dropouts / Dead Elements Acoustic Lens Damage Cable Damage Connector Damage Weak or Noisy Image If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do
L12-5 linear array ultrasound probe for repair
L12-5 linear array ultrasound probe for repair ► Brief Description Fault: Crystal damage. Solution: Replace crystal. ►Problem & Solutoin Model L12-5 Fault Crystal damage Solution Replace crystal ► Common issues with L12-5 Probe Image Dropouts / Dead Elements Acoustic Lens Damage Cable Damage Connector Damage Noisy or Attenuated Image If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve
EPlQ/X5-1 Probe 2D Image Abnormal For Repair
EPlQ/X5-1 Probe For Repair ► Brief Description Fault: 2D image abnormal Solution: Replace scanhead ►Problem & Solutoin Model EPlQ/X5-1 Fault 2D image abnormal Solution Replace scanhead ► Common Problems & Symptoms Image Artifacts (black lines, dropouts, flickering) → Damaged elements or cable No Image/Dead Probe → Cable break, connector issues, or internal failure Lens Damage (cracks, delamination) → Acoustic coupling issues Error Messages ("Probe Not Recognized") → Faulty
C5-2 Convex Probe For Repair Medical Device Parts
C5-2 Convex Probe For Repair ► Brief Description Fault: Strain relief damage, cable broken. Solution: Replace strain relief, repair cable. ►Problem & Solutoin Model C5-2 Convex Fault Strain relief damage, cable broken Solution Replace strain relief, repair cable ► Common Issues with C5-2 Probes Dead or flickering elements (image dropouts, black lines) Cable damage (broken wires near the connector or handle) Lens damage (cracks, delamination, or wear) Acoustic array failure
1898200T M4 Straightshot Microdebrider and Handpiece for repair
1898200T M4 Straightshot Microdebrider and Handpiece for repair ► Brief Description Fault: 1) The bearing is corroded and the seal is broken. 2) Motor stuck, inside is full of disinfection residue, motor corrosion. 3) The rotation direction of the knife head is stuck. 4) Handle paint off, internal corrosion.. Solution: Exchange the whole handpiece except cable. ►Problem & Solutoin Model 1898200T Fault 1) The bearing is corroded and the seal is broken. 2) Motor stuck, inside
Samsung Medison 3D4-9 Probe SN P0N7 For Repair
Samsung Medison 3D4-9 Probe SN P0N7 For Repair ► Brief Description Fault: Scanhead broken. Solution: Replace scanhead ► Problem & Solutoin Brand Samsung Model 3D4-9 Fault Scanhead broken Solution Replace scanhead ► Common i ssues with 3D4-9 Probes Cable damage Lens delamination Element failure (3D/4D imaging issues) Connector problems If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you
Medical Device Repair Cases LINVATEC HD 4300 Arthroscope For Repair
LINVATEC HD 4300 Arthroscope for repair, SN: 62**** Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Replace objective lens, inner optical parts, clean and adjust optical system, seal with eyepiece. Image Quality Related Issues: 1.Fogging or Blurry Vision: This is a frequent problem in arthroscopy. It can be due to condensation inside the endoscope, debris on the lens, or issues with the camera system. 2.Dark or Dim Image: This could be
Medical Device Repair Cases Stryker 30º 0502-503-030 Laproscope For Repair
Stryker 30º 0502-503-030 laproscope, SN:103**** Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. Repair cost: $*** Common Issues with the Stryker 0502-503-030 Laparoscope: 1.Fogging/Obscured Vision: This is a frequent issue during laparoscopic procedures. It can be caused by temperature differences between the endoscope and
FUJINON EG-530N gastroscope for repair
FUJINON EG-530N gastroscope for repair ► Brief Description Fault: 1.Biopsy tube broken; 2.Angle is not enough; 3.Camera over-heating, there are occasional snowflakes on the image and big risk for CCD; 4.Insert tube aging and wrinkle; 5.Light guide tube aging and wrinkle; 6.Stress boot of light guide tube broken; 7.Light bundle broken; 8.EVE tube aging and wrinkle. Solution: 1: Replace CCD; insert tube; light bundle; biopsy tube; bending mesh; wires; bending section; rubber;
Karl Storz 27005BA Cystoscope for repair
Karl Storz 27005BA Cystoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope and test. ►Problem & Solutoin
Olympus A4674A Autoclavable Hysteroscope for repair
Olympus A4674A Autoclavable Hysteroscope for repair ► Brief Description Fault: 1. Image blur and black shadow. 2. Rod lens broken. 3. Outer tube bending and transformation. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Clean and install the optical path, check the image. 8. Seal with endoscope and test. ►Problem & Solutoin Brand Olympus Model A4674A Fault 1. Image blur