Product Overview

The ISO 5660 Cone Calorimeter is a precision laboratory instrument designed to measure heat release rate (HRR), smoke production, ignition time, oxygen consumption, and combustion gas generation for building materials and other specimens exposed to controlled radiant heating.

Operating Principle

Based on the oxygen consumption calorimetry principle, the instrument calculates heat release using the established correlation that 13.1 MJ of energy is released per kilogram of oxygen consumed during combustion.

Compliance Standards
  • ISO 5660: Reaction to fire tests - Heat release, smoke production and mass loss rate
  • ASTM E 1354: Standard test method for heat and visible smoke release rates
  • BS 476 Part 15: Heat release rate testing of building materials
  • GB/T 16172-2007: Test method for heat release rate of building materials
  • NFPA 264: Standard method of test for heat and visible smoke release rates
Technical Specifications
Parameter Specification
Heater Power 5000W (0-100kW/m² PID temperature control)
Sample Size ≤100×100×50mm
Gas Analysis Magnetic oxygen (O₂) + NDIR (CO/CO₂)
Smoke Detection Laser photoelectric method (dual detector system)
Data Output HRR, total smoke production, CO/CO₂ yield, mass loss rate
Cooling System Portable water cooling, no external water source required
Instrument Images
ISO 5660 Cone Calorimeter laboratory testing equipment with control panel and sample chamber Close-up view of cone calorimeter heating element and sample positioning system Cone calorimeter data display screen showing real-time measurement parameters and test results
Measured Parameters
  • Heat Flux (kW/m²)
  • Exhaust duct flow rate (l/s)
  • C factor (1/kg * m * K)
  • Ignition time and Extinguishment time (s)
  • Total oxygen consumption (g)
  • Total Smoke Release (m/m²)
  • Mass loss and Mass loss rate (g, g/s)
  • Heat release rate (MJ/kg, kW/m²)
  • Effective heat of combustion (MJ/kg)
  • Carbon monoxide yield (kg/kg)
  • Carbon dioxide yield (kg/kg)
Operation Guidelines
Pre-Test Preparation
  • Follow instruction manual procedures strictly for safety compliance
  • Perform sample pretreatment, cutting, and chamber cleaning before testing
  • Wear appropriate personal protective equipment including protective clothing, gloves, and goggles
During Testing
  • Monitor instrument status and data output regularly
  • Verify measurement accuracy throughout the process
Maintenance Requirements
  • Perform regular cleaning and maintenance to ensure measurement accuracy and instrument longevity
  • Replace consumables including gas cylinders and filters periodically
  • Disassemble and clean instrument components after use
  • Store in dry, well-ventilated environment

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Chongqing gold mechnical and electrical equipment Co., Ltd
Location 25F, China Resources Building, Xiejiawan, Jiulongpo District,Chongqing,400050,China
Contact Person Ivy Zhang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.