Product Overview
The ISO 5660 Cone Calorimeter is a precision laboratory instrument designed to measure heat release rate (HRR), smoke production, ignition time, oxygen consumption, and combustion gas generation for building materials and other specimens exposed to controlled radiant heating.
Operating Principle
Based on the oxygen consumption calorimetry principle, the instrument calculates heat release using the established correlation that 13.1 MJ of energy is released per kilogram of oxygen consumed during combustion.
Compliance Standards
- ISO 5660: Reaction to fire tests - Heat release, smoke production and mass loss rate
- ASTM E 1354: Standard test method for heat and visible smoke release rates
- BS 476 Part 15: Heat release rate testing of building materials
- GB/T 16172-2007: Test method for heat release rate of building materials
- NFPA 264: Standard method of test for heat and visible smoke release rates
Technical Specifications
| Parameter | Specification |
|---|---|
| Heater Power | 5000W (0-100kW/m² PID temperature control) |
| Sample Size | ≤100×100×50mm |
| Gas Analysis | Magnetic oxygen (O₂) + NDIR (CO/CO₂) |
| Smoke Detection | Laser photoelectric method (dual detector system) |
| Data Output | HRR, total smoke production, CO/CO₂ yield, mass loss rate |
| Cooling System | Portable water cooling, no external water source required |
Instrument Images
Measured Parameters
- Heat Flux (kW/m²)
- Exhaust duct flow rate (l/s)
- C factor (1/kg * m * K)
- Ignition time and Extinguishment time (s)
- Total oxygen consumption (g)
- Total Smoke Release (m/m²)
- Mass loss and Mass loss rate (g, g/s)
- Heat release rate (MJ/kg, kW/m²)
- Effective heat of combustion (MJ/kg)
- Carbon monoxide yield (kg/kg)
- Carbon dioxide yield (kg/kg)
Operation Guidelines
Pre-Test Preparation
- Follow instruction manual procedures strictly for safety compliance
- Perform sample pretreatment, cutting, and chamber cleaning before testing
- Wear appropriate personal protective equipment including protective clothing, gloves, and goggles
During Testing
- Monitor instrument status and data output regularly
- Verify measurement accuracy throughout the process
Maintenance Requirements
- Perform regular cleaning and maintenance to ensure measurement accuracy and instrument longevity
- Replace consumables including gas cylinders and filters periodically
- Disassemble and clean instrument components after use
- Store in dry, well-ventilated environment
Related Videos
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.
Company
Chongqing gold mechnical and electrical equipment Co., Ltd
Location
25F, China Resources Building, Xiejiawan, Jiulongpo District,Chongqing,400050,China
Contact Person
Ivy Zhang