Deionized Water pH Sensor with 2-12 pH Range and 0-80℃ Temperature for Pure Water Applications

Price Negotiable
Price: USD50-200
MOQ: 1 Unit
Delivery Time: 5-8 working days
Brand: JUGE
Product Description
Product Overview

Professional deionized water pH sensor designed for pure water applications including RO water and deionized water systems. Features polypropylene housing for chemical resistance and reliable performance in demanding industrial environments.

Technical Specifications
Parameter Specification
pH Range 2-12 pH
Temperature Range 0-80℃
Pressure Resistance 0-0.3 MPa
Housing Material PP (Polypropylene)
Connection Thread NPT 3/4"
Cable Length 5m standard (customizable)
Membrane Resistance <100 MΩ
Measurement Technology

Based on the Nernst equation principle, this potentiometric sensor functions as a galvanic cell converting chemical reaction energy into electrical energy. The system consists of measuring and reference half-cells with Ag/AgCl reference system and PTFE liquid junction for stable performance in pure water applications.

Installation Methods
Deionized water pH sensor installation diagram showing multiple mounting options
  • Sidewall Installation: Ensure interface tilt angle exceeds 15 degrees
  • Top Flange Installation: Note flange dimensions and electrode insertion depth
  • Pipe Installation: Consider pipe diameter, flow velocity, and pressure
  • Flow-Through Installation: Monitor flow velocity and pressure parameters
  • Submersion Installation: Adjust bracket length according to application requirements
Maintenance & Care
  • Unscrew protective cap and immerse electrode bulb and liquid junction in sample before use
  • Rinse with water if salt crystals form near electrode tip
  • Check for air bubbles in glass bulb by shaking electrode vertically
  • Keep glass membrane hydrated for fast response time
  • Store with 3 mol/L KCl solution in protective cap
  • Clean aged electrodes with 10% HCl solution for optimal performance
Product Lifespan & Calibration

Typical lifespan exceeds 12 months with proper maintenance. Shelf life of 12-24 months when stored correctly. Regular calibration using fresh pH/ORP standard buffer solutions ensures measurement accuracy throughout the product lifecycle.

Note: Electrodes are consumable components and generally not covered by warranty. Exposure to strong acids, bases, high temperatures, or corrosive liquids may shorten operational lifespan.

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Nanjing Juge Environmental Technology Co., Ltd.
Location Building 17, No. 99 Guangli Road, Moling Subdistrict, Jiangning District, Nanjing, Jiangsu
Contact Person Leon Chen

Request A Quote

Please check your email address.
Your message must be at least 20 characters.