Online COD Analyzer CODmn Water Quality Monitor with 0-20 mg/L Range and 1500mm Height for Wastewater Analysis
Price:
USD2000-7000
MOQ:
1 Unit
Delivery Time:
5-8 working days
Brand:
JUGE
Product Description
Online COD Analyzer for Water Quality Monitoring
Professional permanganate index analyzer designed for wastewater treatment applications with measurement range of 0-20 mg/L.
Technical Specifications
| Parameter | Specification |
|---|---|
| Measurement Method | Potassium permanganate oxidation spectrophotometry |
| Measurement Range | 0-20 mg/L |
| Accuracy | ≤ ±10% |
| Repeatability | ≤ ±5% |
| Measurement Cycle | Minimum 20 minutes digestion, adjustable 5-120 minutes |
| Sampling Cycle | Interval measurement (20-9999 minutes) and hourly modes |
| Calibration Cycle | Adjustable 1-99 days |
| Maintenance Cycle | Monthly, approximately 30 minutes |
| Reagent Consumption | < ¥0.5 per sample |
| Output Interfaces | RS-232, 4-20mA, RS-485 |
| Environmental Requirements | +5-28°C temperature, ≤90% humidity (non-condensing) |
| Power Supply | AC220V±10%, 50Hz±10%, 5A |
| Dimensions | 1500mm(H) × 550mm(W) × 450mm(D) |
Key Features & Design Advantages
Selector Valve Assembly
Flexible reagent sampling timing and diverse channel configurations with minimized dead volume for easy maintenance and extended service life.
Micro-Dosing Unit
Visual photoelectric system ensures precise reagent measurement with 0.7 mL per dose, eliminating errors from pump tube wear.
Sample Introduction Unit
Peristaltic pump with negative pressure suction creates air buffer zone to prevent corrosion and enhance reagent mixing flexibility.
Sealed Digestion Unit
High-temperature, high-pressure operation accelerates reactions while eliminating corrosion risks from volatile gases.
Reagent Tubing System
Imported modified PTFE transparent tubing with inner diameter >1.5 mm minimizes clogging from water sample particulates.
Analytical Principle
The water sample is mixed with potassium permanganate digestion solution and dilute sulfuric acid, then heated for digestion. After permanganate ions oxidize organic matter, the resulting color change is detected and converted into permanganate index values for output.
Additional Capabilities
- Data retention during abnormal alarms and power outages
- Touchscreen display and command input interface
- Automatic discharge of residual reactants and operational resumption after reset or power restoration
Related Videos