PVC Membrane Potassium Ion Sensor with 0.04-39,000 mg/L Range and 5m Cable for Industrial Water Analysis

Price Negotiable
Price: USD300-600
MOQ: 1 Unit
Delivery Time: 5-8 working days
Brand: JUGE
Product Description
Technical Specifications
ParameterSpecification
Sensor TypePVC Membrane Ion-Selective Electrode
Measurement Range0.04-39,000 mg/L
pH Range2-12 pH
Temperature Range0-50°C
Pressure ResistanceNot pressure-resistant
Temperature SensorNone
Housing MaterialEP (Engineering Plastic)
Connection ThreadPG13.5
Cable Length5m standard (customizable)
Cable ConnectorTinned standard (customizable)
Product Overview

This PVC membrane ion-selective electrode (ISE) is designed for precise potassium ion concentration measurement in aqueous solutions. Utilizing membrane potential technology, the sensor provides accurate readings by detecting ion activity relative to a reference electrode, enabling reliable quantitative analysis for industrial and laboratory applications.

Key Features
  • Reagent-free operation eliminates chemical waste and reduces operational costs
  • Fast response time with stable signal output using redox potential measurement
  • Durable polypropylene electrode construction ensures long service life
  • High-quality shielded cables maintain signal accuracy and stability
  • Environmentally friendly design with no pollution emissions
Pre-Measurement Preparation
  1. Remove protective cap from electrode tip - avoid touching sensitive membrane
  2. Shake electrode downward to remove internal air bubbles
  3. For new or unused electrodes: activate by soaking in 1 ppm standard solution for 24 hours
  4. Rinse with deionized water and blot dry with non-woven cloth
  5. Before testing: immerse in deionized water for 5 minutes with stirring
Calibration Procedure
  1. Connect electrode to meter and set to mV mode
  2. Prepare 100 mL deionized water in 150 mL beaker
  3. Immerse dried electrode in deionized water
  4. Add 1 mL of 1000 ppm standard solution, stir, record reading (E₁)
  5. Add 10 mL of same standard solution, stir, record reading (E₂)
  6. Calculate slope at 25°C as E₂ - E₁
Storage Guidelines

Short-term: Preserve in deionized water or 1 ppm standard solution

Long-term: Rinse, dry thoroughly, and store in original packaging

OEM & ODM Services

Customizable cable lengths, connectors, and sensor configurations available to meet specific application requirements and integration needs.

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Nanjing Juge Environmental Technology Co., Ltd.
Location Building 17, No. 99 Guangli Road, Moling Subdistrict, Jiangning District, Nanjing, Jiangsu
Contact Person Leon Chen

Request A Quote

Please check your email address.
Your message must be at least 20 characters.