Imported PVC Membrane Ammonium Ion Selective Electrode ISE Sensor with 0.02-18000 mg/L Range and 10m Cable
Price:
USD300-600
MOQ:
1 Unit
Delivery Time:
5-8 working days
Brand:
JUGE
Product Description
Technical Specifications
| Parameter | Specification |
|---|---|
| Measurement Range | 0.02-18000 mg/L |
| pH Range | 4-10 pH |
| Temperature Range | 0-50°C |
| Pressure Resistance | Non-pressure resistant |
| Temperature Sensor | NTC10K/NTC2.2/PT100/PT1000 |
| Housing Material | PP (Polypropylene) |
| Connection Thread | NPT3/4'' |
| Cable Length | 10m standard (customizable) |
| Cable Termination | Tinned standard (customizable) |
Product Overview
The imported PVC membrane ammonium ion-selective electrode (ISE) is a precision sensor that measures NH4+ ion concentration in solutions. This membrane electrode generates electrical potential through ion exchange or diffusion processes, enabling accurate quantitative analysis of ammonium ion activity relative to concentration levels.
Key Features
- Reagent-free operation eliminates chemical waste and reduces environmental impact
- Fast response time with stable signal output using redox potential measurement
- Durable PP (polypropylene) housing material for long service life
- High-quality sensor-specific cables ensure signal accuracy and stability
- Cost-effective and eco-friendly analytical solution
Pre-Measurement Preparation
- Remove protective cap from electrode tip - avoid touching sensitive membrane
- Shake electrode downward to remove internal air bubbles
- Activate new or long-unused electrodes by soaking in 1 ppm standard solution for minimum 24 hours
- After activation, rinse with deionized water and gently blot dry
- Before testing, immerse electrode tip in deionized water for 5 minutes with stirring
Replace rinse water multiple times and repeat rinsing to prevent measurement errors
Calibration Procedure
- Connect electrode to meter and set to mV mode
- Pour 100 mL deionized water into 150 mL beaker and stir thoroughly
- Blot electrode dry and immerse in prepared deionized water
- Add 1 mL of 1000 ppm standard solution, stir, record stable reading (E₁)
- Add 10 mL of same standard solution, stir, record stable reading (E₂)
- Calculate slope at 25°C as E₂ - E₁ and compare with specifications
Troubleshooting
- If slope deviates from specified range, prepare freshly diluted standard solution
- Soak electrode in diluted solution for 10 minutes before testing
- Repeat calibration steps - if issue persists, contact technical support
Storage Guidelines
- Short-term: Preserve in deionized water or 1 ppm standard solution
- Long-term: Rinse with deionized water, wipe membrane dry, store in original packaging
- Reactivation required after long-term storage: soak in 1 mg/L standard solution for over 2 hours