Imported PVC Membrane Ammonium Ion Selective Electrode ISE Sensor with 0.02-18000 mg/L Range and 10m Cable

Price Negotiable
Price: USD300-600
MOQ: 1 Unit
Delivery Time: 5-8 working days
Brand: JUGE
Product Description
Technical Specifications
ParameterSpecification
Measurement Range0.02-18000 mg/L
pH Range4-10 pH
Temperature Range0-50°C
Pressure ResistanceNon-pressure resistant
Temperature SensorNTC10K/NTC2.2/PT100/PT1000
Housing MaterialPP (Polypropylene)
Connection ThreadNPT3/4''
Cable Length10m standard (customizable)
Cable TerminationTinned standard (customizable)
Product Overview

The imported PVC membrane ammonium ion-selective electrode (ISE) is a precision sensor that measures NH4+ ion concentration in solutions. This membrane electrode generates electrical potential through ion exchange or diffusion processes, enabling accurate quantitative analysis of ammonium ion activity relative to concentration levels.

Key Features
  • Reagent-free operation eliminates chemical waste and reduces environmental impact
  • Fast response time with stable signal output using redox potential measurement
  • Durable PP (polypropylene) housing material for long service life
  • High-quality sensor-specific cables ensure signal accuracy and stability
  • Cost-effective and eco-friendly analytical solution
Pre-Measurement Preparation
  • Remove protective cap from electrode tip - avoid touching sensitive membrane
  • Shake electrode downward to remove internal air bubbles
  • Activate new or long-unused electrodes by soaking in 1 ppm standard solution for minimum 24 hours
  • After activation, rinse with deionized water and gently blot dry
  • Before testing, immerse electrode tip in deionized water for 5 minutes with stirring
Replace rinse water multiple times and repeat rinsing to prevent measurement errors
Calibration Procedure
  • Connect electrode to meter and set to mV mode
  • Pour 100 mL deionized water into 150 mL beaker and stir thoroughly
  • Blot electrode dry and immerse in prepared deionized water
  • Add 1 mL of 1000 ppm standard solution, stir, record stable reading (E₁)
  • Add 10 mL of same standard solution, stir, record stable reading (E₂)
  • Calculate slope at 25°C as E₂ - E₁ and compare with specifications
Troubleshooting
  • If slope deviates from specified range, prepare freshly diluted standard solution
  • Soak electrode in diluted solution for 10 minutes before testing
  • Repeat calibration steps - if issue persists, contact technical support
Storage Guidelines
  • Short-term: Preserve in deionized water or 1 ppm standard solution
  • Long-term: Rinse with deionized water, wipe membrane dry, store in original packaging
  • Reactivation required after long-term storage: soak in 1 mg/L standard solution for over 2 hours

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Nanjing Juge Environmental Technology Co., Ltd.
Location Building 17, No. 99 Guangli Road, Moling Subdistrict, Jiangning District, Nanjing, Jiangsu
Contact Person Leon Chen

Request A Quote

Please check your email address.
Your message must be at least 20 characters.