Large Memory Sodium Ion Analyzer Online Water Analyzer Instrument RS485 Interface
Online Water Quality Hydrazine Analyzer
>>Application
With the increasing regulatory standards in power plants and the rising requirements for demineralized water and condensate treatment processes, the timeliness and accuracy of sodium ion monitoring have become significantly more critical. To address this, our company has developed the AG9264C Online Sodium Ion Analyzer. Featuring a constant-pressure water sample system and continuous alkalization design, this instrument ensures precise and reliable measurements.
>>Features
- Constant-pressure water sample ensures stable alkalization effects.
- Unique flow path system guarantees rapid and accurate measurements.
- Built-in calibration pump simplifies calibration procedures.
- 2–5-point calibration enhances measurement accuracy.
- Large-capacity memory stores measurement data for up to 5 years.
- 7-inch touch color LCD screen provides intuitive operation and clear visualization.
- 2 isolated current outputs (configurable for any range, linear or logarithmic output).
- 2 relay outputs (configurable as high/low limit alarms).
- RS485 interface enables remote data monitoring.
- Queryable data for curves and historical records within any timeframe.
- Wide measurement range: pNa2–pNa9 (229,000–0.23 μg/L).
- Uncertainty: ±0.05 pNa or ±1 μg/L.
>>Technical Parameter
|
Specification |
Detail |
|
Measurement range |
(2.00~9.00)pNa(229000μg/L~0.23μg/L) |
|
Resolution |
0.01pNa or 0.1μg/L |
|
Display error |
±0.05pNa or ±5μg/L |
|
Temperature Sensor |
NTC10K |
|
Temperature measurement range |
(0.0~60.0) ℃ |
|
Temperature measurement accuracy |
± 0.5℃ |
|
Temperature compensation range |
(0.1~60.0) ℃ |
|
Isolate the output signal and load |
4 ~ 20 mA (RL <500 Ω) |
|
Alarm contact capacity |
normally open contact AC220V / 3A, DC24V / 5A |
|
Working conditions |
Temperature:0~45℃ Humidity: ≤85% |
|
Sample conditions |
Flow rate:100-300 mL/min Temperature:(5~45)℃ Pressure:50-300 KPa |
|
Permissible solid composition of water sample |
≤ 5 microns (no glue is allowed, no grease) |
|
Power supply |
AC220V± 10% 50Hz±0.5Hz power: 150W |
|
Instrument external dimensions |
height, 720, width, 460, depth, 360 (mm) |
|
Instrument opening size |
665x410mm |
>>Dimension