Explosion-Proof ORP Meter with 4-20mA HART Output -1000 to +1000 mV Range IP66 Protection
Price:
USD50-200
MOQ:
1 Unit
Delivery Time:
5-8 working days
Brand:
JUGE
Product Description
Industrial Explosion-Proof ORP Monitoring System
This microprocessor-based online ORP meter provides continuous monitoring and control of oxidation-reduction potential and temperature values in aqueous solutions. Designed for demanding industrial environments with explosion-proof certification Exia IICT6 Ga and IP66 protection.
Key Features
- Intelligent menu operation with password protection
- Multiple automatic calibration functions
- Manual and automatic temperature compensation
- Two-wire system with 4-20mA HART output
- Explosion-proof design for hazardous environments
Technical Specifications
| Parameter | Specification |
|---|---|
| ORP Measurement Range | -1000 to +1000 mV |
| Temperature Range | -20 to 150.0°C |
| ORP Resolution | 1 mV |
| Temperature Resolution | 0.1°C |
| ORP Accuracy | ±2 mV |
| Temperature Accuracy | ±0.3°C |
| Signal Output | 4-20mA analog (load resistance <500Ω) with optional HART protocol |
| Power Supply | 24 VDC, ≤0.3W consumption |
| Protection Rating | IP66, Explosion-proof Exia IICT6 Ga |
| Controller Weight | 0.6 kg |
Measurement Principle
Based on the Nernst equation, this ORP meter uses galvanic cell sensors to convert chemical reaction energy into electrical energy. The system measures oxidation-reduction potential, which reflects the collective redox behavior of all substances in solution, providing critical insights into water quality and system ecology.
Installation Options
- Sidewall Installation: Ensure interface tilt angle exceeds 15 degrees
- Top Flange Installation: Note flange dimensions and electrode insertion depth
- Pipe Installation: Consider pipe diameter, flow velocity, and pressure
- Flow-Through Installation: Monitor flow velocity and pressure parameters
- Submersion Installation: Adjust bracket length as required
Electrode Maintenance
- Before use, unscrew protective cap and immerse electrode bulb in sample
- Rinse with water if salt crystals form near electrode tip
- Check for air bubbles by shaking electrode vertically
- Keep glass membrane hydrated for optimal response time
- Store with 3 mol/L KCl solution in protective cap
- Clean aged electrodes with 10% HCl solution
- Regularly calibrate using fresh pH/ORP standard buffer solutions
Electrode Lifespan: Under normal conditions with proper maintenance, electrodes typically last over 12 months. Even unused electrodes have a shelf life of 12-24 months when stored correctly. Exposure to strong acids, bases, high temperatures, or corrosive liquids may shorten lifespan. Electrodes are consumables and generally not covered by warranty.