Mold incubator with temperature humidity control NovaIncu MD1000-I
- Constructed with premium materials: stainless steel chamber, powder-coated housing, toughened glass.
- Designed for efficiency with an inner glass door and an outer insulated door.
- Compressor system provides consistent cooling with intelligent self-management features.
- Guarantees excellent temperature uniformity through omnidirectional 3D air flow.
- Standard UV illumination supports chamber sanitation between runs.
- Unifies precise control of temperature and automatic humidity in one system.
- Automated water feeding from a dual-tank system prevents humidification failure.
- Dynamic协同 technology optimizes controller interaction for stable conditions.
- Offers complete automation via 99-hour programmable timer and memory functions.
- Safety is enhanced with multiple alarms and an optional independent temperature guardian.
| Model | NovaIncu MD1000-I | |
| Chamber volume(L) | 1000 | |
| Temperature Control Range | 5℃~50℃ | |
| Temperature | Resolution | 0.1℃ |
| Fluctuation | ±0.5℃ | |
| Uniformity | ±2℃ at 37℃ | |
| controller | PID microprocessor control, soft touch, LED display | |
| Sensor | Pt 100 resistor | |
| Timer | Power-on, power off and working. Timing range: 1min-99hr | |
| Humidity range | 50%-90% | |
| Humidity | Accuracy | ±0.1%RH |
| Fluctuation | ±3%RH | |
| controller | PID microprocessor control, soft touch, LED display | |
| Sensor | Capacitor type | |
| Material | Internal | Mirror polished 304 stainless steel |
| External | Steel (powder coating) | |
| Dimensions (WxDxH,cm) | Internal | 95x90x115 |
| External | 110x115x198 | |
| Net Weight(Kg) | 315 | |
| Consumption Power(W) | 4000 | |
| Shelf Size(mm) | 928*880 | |
| Shelf Qty(Standard/Maximum) | 3/18 | |
| Power Supply | 220V/50Hz (Optional: 220V/60Hz, 110V/60Hz) | |
For manufacturers and high-compliance laboratories, incubator quality is governed by formal standards and certifications. Key standards include ISO 9001 for overall quality management systems and specific clauses within ISO/IEC 17025 for testing and calibration laboratories. While there is no single “ISO for incubators," performance is often validated against manufacturer specifications which should be traceable.
In vitro diagnostic (IVD) and pharmaceutical manufacturing require adherence to Good Manufacturing Practice (GMP), where equipment must be qualified (DQ, IQ, OQ, PQ). Data logging features must comply with FDA 21 CFR Part 11 for electronic records and signatures, ensuring audit trails for all parameter changes and alarm events. Independent safety certifications like UL, CE, or ETL are mandatory for electrical safety.
Furthermore, environmental simulation chambers used for stability testing follow ICH Q1A(R2) guidelines for climatic zones. When procuring an incubator for a regulated environment, the documentation package—including detailed calibration certificates, uniformity mapping reports, and software validation documents—is as important as the hardware itself. This framework of quality control ensures that every incubator performs as a validated, reliable component within a system dedicated to producing accurate, safe, and legally defensible results.
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.