Shaanxi Qindiyuan Environmental Protection Equipment Co., Ltd.
                                                                                                           
Verified Supplier
8 Years
Since 2018
Menu

2 Wire HART 7 Temperature Transmitters PR 6437D Single Or True Dual Inputs

Price Negotiable
Price: 200-400usd
MOQ: 1
Delivery Time: 5-7 work days
Brand: PR
Product Description
2-wire HART 7 Temperature Transmitter PR 6437D
Key Features
  • RTD, TC, potentiometer, linear resistance and bipolar mV input compatibility
  • Single or true dual inputs with sensor redundancy and drift detection
  • Wide ambient operating temperature range
  • HART 7 protocol support
Environmental Conditions
Operating temperature -50°C to +85°C (standard) -40°C to +80°C (SIL)
Storage temperature -50°C to +85°C
Calibration temperature 23...25°C
Relative humidity < 99% RH (non-cond.)
Protection degree IP20
Mechanical Specifications
Dimensions (HxWxD) 109 x 23.5 x 104 mm
Weight (single input / dual input) 150 g / 160 g
Wire size 0.13...2.08 mm² AWG 26...14 stranded wire
DIN rail type DIN EN 60715/35 mm
Screw terminal torque 0.5 Nm
Vibration IEC 60068-2-6
2...25 Hz: ±1.6 mm
25...100 Hz: ±4 g
Common Specifications
Supply
Supply voltage 7.5*...30** VDC
Power dissipation, per channel ≤ 850 mW
Additional min. supply voltage when using test terminals 0.8 V
Min. load resistance at >37 V supply (Vsupply - 37) / 23 mA
Performance Characteristics
Isolation voltage, test / working 2.5 kVAC / 42 VAC
Response time 75 ms
Programmable damping 0...60 s
Warm-up time < 5 min.
Start-up time < 2.75 s
Long-term stability ±0.05% of span / year (±0.18% of span / 5 years)
Signal dynamics, input / output 24 bit / 18 bit
Input Specifications
RTD Input
RTD type Pt10...10000, Ni10...10000, Cu5...1000
Basic accuracy, e.g. Pt100 ≤ ±0.04°C
Cable resistance per wire 50 Ω (max.)
TC Input
Thermocouple type B, E, J, K, L, N, R, S, T, U, W3, W5, LR
Basic accuracy, e.g. TC K ≤ ±0.25°C
Output Specifications
Normal range, programmable 3.8...20.5 / 20.5...3.8 mA
Extended range (output limits), programmable 3.5...23 / 23...3.5 mA
Basic accuracy ≤ ±1.6 μA (0.01% of full output span)
HART protocol revisions HART 7 and HART 5
Certifications and Approvals
ATEX DEKRA 16ATEX0047X
IECEx IECEx DEK. 16.0029X
CSA CSA 16.70066266
SIL SIL 2 / SIL 3 certified & fully assessed acc. to IEC 61508
* / ** See manual for details
Similar Products

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.