China S5-2 Cardiac probe for repair for sale

S5-2 Cardiac probe for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

S5-2 Cardiac probe for repair Fault: Lens problem, cable broken, without strain relief. Solution: Replace probe lens, repair cable, replace strain relief. ►Problem & Solutoin Model S5-2 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, repair cable, replace strain relief ► Common issues with S5-2 Cardiac probe Poor or blurry images Constant noise or static Color streaking or intermittent artifacts Cracked or dried-out lens Cable or strain

China L12-3 Ultrasuound probe for repair for sale

L12-3 Ultrasuound probe for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

L12-3 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Model L12-3 Fault Lens problem Solution Replace lens ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket Damage Connector Damage Strain Relief Separation Membrane/Lens Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue.

China L12-3 linear Probe for repair for sale

L12-3 linear Probe for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

L12-3 linear Probe for repair ► Brief Description Fault: Array element aging, lens bubble. Solution: Replace Crystal. ►Problem & Solutoin Model L12-3 Fault Array element aging, lens bubble Solution Replace Crystal ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket Damage Connector Damage Strain Relief Separation Membrane/Lens Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will

China EPlQ/X5-1 Probe 2D Image Abnormal For Repair for sale

EPlQ/X5-1 Probe 2D Image Abnormal For Repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

EPlQ/X5-1 Probe For Repair ► Brief Description Fault: 2D image abnormal Solution: Replace scanhead ►Problem & Solutoin Model EPlQ/X5-1 Fault 2D image abnormal Solution Replace scanhead ► Common Problems & Symptoms Image Artifacts (black lines, dropouts, flickering) → Damaged elements or cable No Image/Dead Probe → Cable break, connector issues, or internal failure Lens Damage (cracks, delamination) → Acoustic coupling issues Error Messages ("Probe Not Recognized") → Faulty

China Samsung Medison 3D4-9 Probe SN P0N7  For Repair for sale

Samsung Medison 3D4-9 Probe SN P0N7 For Repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Samsung Medison 3D4-9 Probe SN P0N7 For Repair ► Brief Description Fault: Scanhead broken. Solution: Replace scanhead ► Problem & Solutoin Brand Samsung Model 3D4-9 Fault Scanhead broken Solution Replace scanhead ► Common i ssues with 3D4-9 Probes Cable damage Lens delamination Element failure (3D/4D imaging issues) Connector problems If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you

China Medical Device Repair Cases LINVATEC HD 4300 Arthroscope For Repair for sale Video

Medical Device Repair Cases LINVATEC HD 4300 Arthroscope For Repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

LINVATEC HD 4300 Arthroscope for repair, SN: 62**** Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Replace objective lens, inner optical parts, clean and adjust optical system, seal with eyepiece. Image Quality Related Issues: 1.Fogging or Blurry Vision: This is a frequent problem in arthroscopy. It can be due to condensation inside the endoscope, debris on the lens, or issues with the camera system. 2.Dark or Dim Image: This could be

China Medical Device Repair Cases Stryker 30º 0502-503-030 Laproscope For Repair for sale Video

Medical Device Repair Cases Stryker 30º 0502-503-030 Laproscope For Repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Stryker 30º 0502-503-030 laproscope, SN:103**** Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. Repair cost: $*** Common Issues with the Stryker 0502-503-030 Laparoscope: 1.Fogging/Obscured Vision: This is a frequent issue during laparoscopic procedures. It can be caused by temperature differences between the endoscope and

China Karl Storz 27005BA Cystoscope for repair for sale

Karl Storz 27005BA Cystoscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Karl Storz 27005BA Cystoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope and test. ►Problem & Solutoin

China Olympus A4674A  Autoclavable Hysteroscope for repair for sale

Olympus A4674A Autoclavable Hysteroscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Olympus A4674A Autoclavable Hysteroscope for repair ► Brief Description Fault: 1. Image blur and black shadow. 2. Rod lens broken. 3. Outer tube bending and transformation. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Clean and install the optical path, check the image. 8. Seal with endoscope and test. ►Problem & Solutoin Brand Olympus Model A4674A Fault 1. Image blur

China LINVATEC HD 4300 Arthroscope for repair for sale

LINVATEC HD 4300 Arthroscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

LINVATEC HD 4300 Arthroscope for repair ► Brief Description Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Solution: Replace objective lens, inner optical parts, clean and adjust optical system. ►Problem & Solutoin Brand LINVATEC Model HD 4300 Fault Image with black shadow; eyepiece with scratches, inner optical parts broken. Solution Replace objective lens, inner optical parts, clean and adjust optical system. ► Common issues with HD

China Stryker 30º 0502-503-030 laproscope for repair for sale

Stryker 30º 0502-503-030 laproscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Stryker 30º 0502-503-030 laproscope for repair ► Brief Description Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. ►Problem & Solutoin Brand Stryker Model 0502-503-030 Fault Image blur, objective lens loose without glue, inner optical parts broken. Solution Add glue for objective lens, replace inner optical parts, clean and

China Olympus A70940A  Autoclavable Arthroscope for repair for sale

Olympus A70940A Autoclavable Arthroscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Olympus A70940A Autoclavable Arthroscope for repair ► Brief Description Fault: 1. Image blur, black dots and dark. 2. Fiber not light enough. 3. Rod lens broken. 4. Objective lens broken and dirty. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Clean and install the optical path, check the image. 8. Seal with endoscope and test. ►Problem & Solutoin Brand Olympus Model

China Olympus GIF-Q150 Gastroscope for repair for sale

Olympus GIF-Q150 Gastroscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Olympus GIF-Q150 Gastroscope for repair ► Brief Description Fault: 1. With image; 2. Light bundlebreak badly; 3. Insert tubeaging; 4. Biopsy tubeabrasion, leakage badly; 5. Angle tied, bendingsection, wires and helix tube abrasion; 6. Rubberwithout glue. Solution: Replace light bundle, insert tube, biopsy tube, bending section, wires, helixtube and rubber. (No warranty for CCD) ►Problem & Solutoin Brand Olympus Model GIF-Q150 Fault 1. With image; 2. Light bundlebreak badly; 3

China GIF-XQ240 Gastroscope by Olympus for repair for sale

GIF-XQ240 Gastroscope by Olympus for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

GIF-XQ240 Gastroscope by Olympus for repair ► Brief Description Fault: Leakage. Solution: Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ►Problem & Solutoin Brand Olympus Model GIF-XQ240 Fault Leakage. Solution Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ► Common issues with GIF-XQ240 Gastroscope No Image / Dark Image Poor Image Resolution / Blurry Image Ineffective Suction Difficulty with Angulation (Stiff or Loose) Bubbles in the Field of View

China Olympus CYF V2 Cystoscope For Repair for sale

Olympus CYF V2 Cystoscope For Repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Olympus CYF V2 Cystoscope For Repair ► Brief Description Fault: Leakage. Solution: Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber. ►Problem & Solutoin Brand Olympus Model CYF V2 Fault Leakage Solution Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber ► Common issues with CYF V2 Cystoscope No Image / Dark Image Poor Image Resolution / Blurry Image Ineffective Suction Difficulty

China Olympus CF-240AI Colonoscope for repair for sale

Olympus CF-240AI Colonoscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Olympus CF-240AI Colonoscope for repair ► Brief Description Fault: Image blur,insert tube aging,light guide squashed,light guide broken 50%,nozzle blocked,angle tight,forceps channel aging. Solution: Replace insertion tube,CCD chips,light guide,forceps channel,water channel,angulation wire,bending mesh,spiral tube,light guide tube,bending rubber. ►Problem & Solutoin Brand Olympus Model CF-240AI Fault Image blur,insert tube aging,light guide squashed,light guide broken 50%

China R Wolf 7326071 ureteroscope for repair for sale

R Wolf 7326071 ureteroscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

R Wolf 7326071 ureteroscope for repair ► Brief Description Fault: The image has spots,image bundle aging,light bundle broken;Insert tube aging;Biopsy channel worn and leakage;Bending section,wire and helix tube were worn. Solution: Replace the light bundle,image bundle,rubber,biopsy channel,helix tube,wire. ►Problem & Solutoin Brand R Wolf Model 7326071 Fault The image has spots,image bundle aging,light bundle broken;Insert tube aging; Biopsy channel worn and leakage;Bending

China Insighters IS3-F laryngoscope for repair for sale

Insighters IS3-F laryngoscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Insighters IS3-F laryngoscope for repair ► Brief Description Fault: 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution: Replace CCD, insert tube, bending section, rubber ►Problem & Solutoin Brand Insighters Model IS3-F Fault 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution Replace CCD, insert tube, bending section, rubber ► Common

China KARL STORZ 11278A1 ureteroscope for repair for sale

KARL STORZ 11278A1 ureteroscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

KARL STORZ 11278A1 ureteroscope for repair ► Brief Description Fault: 1) Image spotted with partially black point, image bundle aged 2) Light bundle partially wire broken 3) Insert tube aged 4) Biopsy tube abrasion and leakage 5) Wires and coil pipe abrasion Solution: Replace HD image bundle, light bundle, insert tube, wires, coil pipe and rubber. ►Problem & Solutoin Brand KARL STORZ Model 11278A1 Fault 1) Image spotted with partially black point, image bundle aged 2) Light

China Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair for sale

Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair

Price: Negotiable
MOQ: 1
Delivery Time: Negotiable

Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair ► Brief Description Fault: 1. Image with black point and image bundle broken 2. Light bundle aged 3. Insert tube aging and wrinkle 4. Bending section deformation 5. Angle with not enough 6. Rubber abrasion Solution: Replace image bundle; light bundle; insert tube; wires; helix tube; bending section; rubber ►Problem & Solutoin Brand Olympus Model ENF-GP Fault 1. Image with black point and image bundle broken 2. Light