Medical Device Repair Cases
Olympus TJF-160VR endoscope for repair
Olympus TJF-160VR endoscope for repair ► Brief Description Fault: 1.No image, CCD broken. 2. Light bundle broken completely. 3. Insert tube aging and wrinkle. 4. Biopsy tube broken. 5. Wires and helix tube broken. 6. clamp wires broken. 7. Operation part broken, no angle. 8. Grip broken. 9. The scope be separated by someone before, some parts missing. Solution: Replace digital parts,CCD,light bundle,insert tube, biopsy tube,air&water tube, bending section, wires,helix tube,
Olympus GIF-Q150 gastroscope for repair
Olympus GIF-Q150 gastroscope for repair ► Brief Description Fault: 1.with image; 2.one eye of light bundle broken; 3.insert tube little aging; 4.biopsy tube abrasion; 5.bending section deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace light bundle ,insert tube, biopsy tube,bending section,wires,helix tube,light guide tube and rubber. ►Problem & Solutoin Brand OLYMPUS Model GIF-Q150 Fault 1.with image; 2.one eye of light
Samsung Medison P2-4AH Cardiac probe for repair
Samsung Medison P2-4AH Cardiac probe for repair ► Brief Description Fault:Lens problem, crystal damage. Solution:Replace probe lens, replace crystal. ►Problem & Solutoin Brand Samsung Model P2-4AH Fault Lens problem, crystal damage Solution Replace probe lens, replace crystal ► Common issues with P2-4AH probe Image Dropout (Dead Elements) Cable Damage Connector Damage Acoustic Lens/Membrane Damage If your instrument has any of the above problems or others, please feel free to
OLYMPUS TJF-Q180V DUODENOSCOPE FOR REPAIR
OLYMPUS TJF-Q180V DUODENOSCOPE FOR REPAIR ► Brief Description Fault: 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. Solution: Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and rubber. (Nowarranty for CCD). ►Problem & Solutoin Brand OLYMPUS Model TJF-Q180V Fault 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5.
Olympus CF-240L Colonoscope for repair
Olympus CF-240L Colonoscope for repair ► Brief Description Fault: Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch. Solution: Replace rubber, objective lens, C cover. Adjust the angle. ►Problem & Solutoin Brand Olympus Model CF-240L Fault Leakage, glass of light bundle blur, rubber leak, angle not enough, image fuzzle,objective lens scratch, Solution Replace rubber, objective lens, C cover. Adjust the angle. ► Common
Karl Storz 27005BA endoscope for repair
Karl Storz 27005BA endoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope and test. ►Problem & Solutoin
Toshiba PST-25BT Phased Array Probe For Repair
Toshiba PST-25BT Phased Array Probe For Repair ► Brief Description Fault: Housing break, image weak, crystal aging. Solution: Replace crystal and housing. ►Problem & Solutoin Brand Toshiba Model PST-25BT Fault Housing break, image weak, crystal aging Solution Replace crystal and housing ► Common issues with PST-25BT Phased Array Probe Image Quality Issues Cable and Connector Issues Probe Casing or Acoustic Lens Damage Temperature Abnormalities Host System Cannot Recognize
STRYKER 1188-210-105 camera head for repair
STRYKER 1188-210-105 camera head for repair ► Brief Description Fault: Image with interrupt. Cable aging. Solution: Replace compatible new cable and connector. ►Problem & Solutoin Brand STRYKER Model 1188-210-105 Fault Image with interrupt. Cable aging Solution Replace compatible new cable and connector ► Common issues with 1188-210-105 camera head Image Quality Problems Physical and Cable Damage Button and Control Malfunctions Overheating If your instrument has any of the
STORZ Camera Head H3-ZA For Repair
STORZ Camera Head H3-ZA For Repair ► Brief Description Fault:Image streaks frequently,cable loose contact and aged. Solution: Replace cable kit. ►Problem & Solutoin Brand STORZ Model H3-ZA Fault Image streaks frequently,cable loose contact and aged Solution Replace cable kit ► Common issues with H3-ZA Camera Image Quality Problems Connection and Communication Issues Functional Abnormalities Physical Damage If your instrument has any of the above problems or others, please
STORZ 13820pks Gastroscope for Repair
STORZ 13820pks Gastroscope for Repair ► Brief Description Fault: 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of. Solution: 1. Replace the wires, coil pipe, rubber; 2.Adjust angle. ►Problem & Solutoin Brand STORZ Model 13820pks Fault 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of Solution 1. Replace the wires, coil pipe, rubber; 2.Adjust angle ► Common issues with 13820pks Gastroscope Dim or Uneven Illumination Fuzzy, Blurry, or
Olympus CF-H190L Colonoscope for repair
Olympus CF-H190L Colonoscope for repair ► Brief Description Fault: 1. Image ok; 2. Light bundle still ok; 3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking; 4. Biopsy tube wear; 5. Left Angle failure, bending section, wire and helix tube wear; 6. Soft and hard adjustment wire wear, relatively tight; 7. Rubber leakage. Solution: Replace the insertion tube, biopsy tube, bending section, wire, helix tube, soft and hard
GE RIC6-12-D endocavity probe for repair
GE RIC6-12-D endocavity probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens,Clean cable. ►Problem & Solutoin Brand GE Model RIC6-12-D Fault Lens problem Solution Replace lens,Clean cable ► Common issues with RIC6-12-D Probe Dome Damage and Fluid Leaks Image Dropouts (Dead Elements) Cable and Connector Issues Motor/Steering Failure (for 4D functionality) If your instrument has any of the above problems or others, please feel free to contact us for
Hitachi C41V1 Ultrasuound probe for repair
Hitachi C41V1 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Brand Hitachi Model C41V1 Fault Lens problem Solution Replace lens ► Common issues with C41V1 Probe Image Dropouts or Streaks Cable and Connector Damage Acoustic Lens Issues Housing Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue. ► About
Sonoscape 8-L743 linear probe for repair
Sonoscape 8-L743 linear probe for repair ► Brief Description Fault: Crystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model 8-L743 Fault Crystal problem Solution Replace the crystal ► Common issues with 8-L743 Probe Acoustic Lens Damage Piezoelectric Crystal Damage (Dead Elements) Cable and Connector Issues Cracked Housing If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our
Sonoscape C362 Abdominal probe for repair
Sonoscape C362 Abdominal probe for repair ► Brief Description Fault: Cystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model C362 Fault Cystal problem Solution Replace the crystal ► Common issues with C362 Probe Image Dropouts (Black Lines or Streaks) Damaged Lens Cracked Housing or Assembly Gaps Cable Issues Color Artifacts in the Image If your instrument has any of the above problems or others, please feel free to contact us for assistance.
GE 4C-RS abdominal probe for repair
GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS Probe Image dropouts / Black lines Acoustic lens damage Cable damage Connector damage Oil leakage If your instrument has any of the above problems or others, please feel free to
Medical Equipment Parts Sonoscape L743 Linear Probe For Repair
Sonoscape L743 linear probe for repair ► Brief Description Fault: Probe head broken. Solution: Replace probe head.(including crystal and upper housing). ►Problem & Solutoin Brand Sonoscape Model L743 Fault Probe head broken Solution Replace probe head.(including crystal and upper housing) ► Common issues with L743 Probes No Image or Intermittent Image Poor Image Quality Probe Not Recognized/Detected Physical Damage to the Probe Head or Cable Overheating Probe Gel Leakage If
Careutech BK Medical 9048 Ultrasound Probe For Repair
BK Medical 9048 Ultrasound Probe For Repair ► Brief Description Fault: The housing break. Solution: Replace probe housings. ►Problem & Solutoin Brand BK Medical Model 9048 Fault The housing break. Solution Replace probe housings. ► Common issues with 9048 Probes No Image / Poor Image Quality Probe Not Recognized by System Intermittent Connection Physical Damage Overheating If your instrument has any of the above problems or others, please feel free to contact us for
BK Medical 9096 Robotic Ultrasound Probe For Repair
BK Medical 9096 Robotic Ultrasound Probe For Repair ► Brief Description Fault: All the outer skin of the cable is seriously damaged. Solution: Replace the outer skin of the cable. ►Problem & Solutoin Brand BK Medical Model 9096 Fault All the outer skin of the cable is seriously damaged Solution Replace the outer skin of the cable ► Common issues with 9096 Probes Cable Damage Acoustic Lens/Membrane Damage Probe Housing Damage Crystal Array Damage Robotic Grasping Interface
OLYMPUS OTV-S6H-1D Camera Head For Repair
Olympus OTV-S6H-1D Camera Head For Repair ► Brief Description Fault: 1.Image yes. 2.Cable aging and interference Solution: Replace new cable. ► Problem & Solutoin Brand Olympus Model OTV-S6H-1D Fault Cable aging and interference Solution Replace cable ► Applications Laparoscopy (abdominal surgery) Arthroscopy (joint surgery) ENT (Ear, Nose, Throat) procedures General endoscopic imaging . ► Troubleshooting & Support If the camera is not functioning, check: Cable connections