Particle filtration efficiency tester(PFE)

 

Scope of application:

It is used to detect the filtration efficiency of particulate matter such as daily masks and filters.

It is applicable to medical equipment inspection center, safety protection inspection center, labor protection inspection center, drug inspection center, disease prevention and control center, hospital, mask and respirator manufacturer, etc.

 

Standards:

YY/T 2626-2019 Respiratory Protection Products

YY0469-2011 Medical Surgical Mask

GB / T 32610-2016 Technical Specifications for Daily Protective Masks

GB / T19083-2010 Technical Requirements for Medical Protective Masks

standard-SC-FT-1702D

NIOSH 42 CFR 84

 

Characteristics :
1. Using a cold aerosol generator to produce continuous and stable aerosol particles, convenient for filling the solution.

2. Use high-precision PM2.5 sensor to measure aerosol concentration.

3. The standard head mould is adopted to imitate the situation when a person wears a protective mask.

4, using a breathing simulator to simulate human breathing.

5. Anti-leakage design of the whole particle to protect the safety of laboratory personnel.

6. The test chamber is equipped with operating gloves to facilitate the replacement of the masks to be inspected.

7, three types of head mold compatibility design, head mold easy to replace.

8. Compatible with mask and respirator samples to test the particle protection and filtering efficiency.

9, aerosol type: salt particulate aerosol and oil particulate aerosol, sodium chloride (optional DOP, DEHS, paraffin oil, corn oil).

 

Technical Parameters:

Flow range

(20 ~ 100) L / min, accuracy 2%

Filtration efficiency

(0.001 ~ 99.999)%

Resistance

0 ~ 2000Pa, accuracy 2%

Test area

100cm²

Aerosol concentration

(12 ~ 20) mg / m³ (salinity), (50 ~ 200) mg / m³ (oily)

 

- Test mode: initial efficiency and loading efficiency

- Test temperature: (25 ± 5) ℃

- Relative humidity: (30 ± 10)%

- External dimensions: 950 × 600 × 600mm

- Power source: AC220V 50Hz

- Gas source: flow 198L / min (0.6Mpa) at 550kPa

- Weight: 100kg

 

-

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company DONGGUAN DAXIAN INSTRUMENT EQUIPMENT CO.,LTD
Location No.168 Zhen'an West Road Changlian Industry Mansion Xiabian Community Dongguan China
Contact Person Sophia Su

Request A Quote

Please check your email address.
Your message must be at least 20 characters.