IEC60529 2013 IPX7 Immersion Test Chamber With Electric Lifting Control Cabinet

Price Negotiable
Price: Customized
MOQ: 1
Delivery Time: 30 Days
Brand: Sinuo
Product Description

 

IEC60529: 2013 IPX7 Immersion Test Chamber With Electric Lifting Control Cabinet

 
Product Overview For IPX7 Immersion Test Chamber:
 
The IPX7 immersion test chamber is designed and manufactured in accordance with the requirements of IEC 60529:2013. Used for testing the IPX7 waterproof level of electrical and electronic products.
The complete set of IPX7 immersion test chamber is composed of water tank, electric lifting basket and control cabinet.
 
Technical parameters For IPX7 Immersion Test Chamber:
 

Chamber external size Approximately W 1080* D 1080*H 2300mm (including casters)
Chamber inner size Length 1000* W 1000* H1800mm
Max. test water depth 1.8m
Observation window Tempered glass (thickness: 12mm)
Observation window width 200mm
Water level adjustment method Water inlet / drain valve, manual adjustment
Water level display Scale (accuracy 1mm)
External chamber material Stainless steel plates + Aluminum profile frame
Inner chamber material 2.0mm thickness stainless steel
Electric lifting basket Made of stainless steel, controlled by electric hoist
Control method Buttons (non-PLC)
Test time 0.01 second-99.99hours can be preset
Immersion depth requirements

The lowest point of enclosure with a height less than 850 mm is located 1000mm below the surface of the water
The highest point of enclosure with a height equal to or greater than 850mm is located 150mm below the surface of the water

Test content IPX7
Power supply AC 220V,50Hz, 1KW
 
Common Troubleshooting For IPX7 Immersion Test Chamber:
 
No.: Failure phenomenon The cause of the failure Approach
1 Equipment is powered on, but timer does not show any value External power supply fails Reconnect power supply
Power control breaker is off Turn on the breaker in the control box
2 No action when the start button is pressed The test time is not cleared Press “Stop/Clear” to clear time
 
 
 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Sinuo Testing Equipment Co. , Limited
Location Room 101, 1st Floor, No. 6, Third Street, Pingshan Industrial Zone 511495, Shibi Street, Panyu District, Guangzhou, China
Contact Person Zoe Zou

Request A Quote

Please check your email address.
Your message must be at least 20 characters.