IEC 60598-1 Clause 8.2.7 Portable Luminaire Residual Voltage Tester
The Residual Voltage Tester complies with clause 8.2.7 of IEC 60598-1:2020 and multiple international standards including IEC 60335-1 Clause 22.5, IEC 60065 Clause 9.1.6, IEC 60601-1, IEC/EN 60598-1 Clause 8.2.7, and IEC 60950-1.
Key Features
- Performs power interruption testing at peak supply voltage for reliable, repeatable measurement results
- Test impedance ≥ 100 MΩ, fully compliant with impedance requirements for residual voltage and capacitance measurements
- Comprehensive testing capabilities including:
- Residual voltage measurement between power terminals
- Residual voltage measurement between power terminals and equipment enclosure
- Residual voltage and capacitance testing of externally accessible capacitors
- Supports both manual and automatic test modes
- Manual mode: Phase-to-phase testing with 10 consecutive measurements per pole combination
- Automatic mode: Sequential three-pole combination tests with 10 measurements per combination
- Automatically records and retains highest measured residual voltage value
- Triggers alarm when residual voltage exceeds 60 V or 34 V
Technical Parameters
| Parameter | Specification |
|---|---|
| Output voltage | AC 0~242V |
| Residual voltage testing range | 5~150V |
| Testing time | 1s |
| Testing impedance | 100 MΩ |
| Testing alarm range | 60/34V |
| Output power | 2000W |
| Testing discrimination | Audible and visual alarm, simultaneously cutting off voltage |
IEC 60598-1 Clause 8.2.7 Requirements
- Luminaires (with exceptions) incorporating capacitors exceeding 0.5 μF must include a discharge device ensuring voltage across the capacitor does not exceed 50 V within 1 minute after disconnection at rated voltage
- Portable luminaires connected via plug, track adaptors, or luminaires with accessible supply connectors containing capacitors exceeding 0.1 μF (or 0.25 μF for luminaires rated below 150 V) must discharge to ≤34 V within 1 second after disconnection
- Other plug-connected luminaires with capacitors exceeding 0.1 μF (or 0.25 μF for luminaires rated below 150 V) and track adaptors must discharge to ≤60 V RMS within 5 seconds after disconnection
- Testing must be conducted with the lamp in circuit when measuring voltage from compensation capacitors if it leads to more onerous results
- Residual voltage measurements are performed on individual luminaires, even when installed in multiple luminaire systems