IEC 62368-1 Clause 5.4.3 CTI PTI Tracking Flammability Test Chamber PLC Control

 

Product Information:

The Tracking Flammability Test Chamber is designed and manufactured according to the standard requirements of IEC 62368-1:2018 clause 5.4.3, IEC 60335-1 clause 29.2, Annex N, IEC 60112, etc.

It is used for determination of the relative resistance to tracking performance that solid electrical insulating material under the action of an electric field, when the surface exposed to water containing impurities.

The working principle of the leakage tracking test (tracking index testing) is that the conducting liquid (0.1%NH 4 CL) of required volume in the required height (35mm) and required time (30s) drops with the voltage between the platinum electrodes (2mm × 5mm) on the surface of solid insulating material. Thus, the users evaluate the tracking resistance performance of solid insulating material surface under the combined influence of electric field and humid or contaminate medium. In a word, this device is used to measure the compare tracking index (CT1) and electrical resistance index (PT1).

 

Technical Parameters:

Applied Standard IEC 62368-1
Clause 5.4.3
Working voltage AC 220V/50Hz,1KVA
Control PLC
Display 7-inch color touch screen
Testing voltage 0~600V adjustable,precision 1.5%
Timing device 9999X0.1S
Electrode Platinum electrode and brass electrode each for a pair
Electrode dimension

(5±0.1)×(2±0.1)×(≥12), 30°slant

Electrode included angle and distance 60°±5°, distance is (4±0.1㎜)
Electrode pressure 1.00N±0.05N(digital display is optional)
The interval of dropping liquid 30±5S, adjustable
Dropping liquid height 35±5mm
Dropping liquid delay 0.2S(solenoid valve opening time)
Dropping liquid size 45~50 drip/cm3
Time-delay circuit 2±0.1S (in 0.5A or larger current)
Short-circuit pressure drop 8% MAX
Short-circuit current 1±0.1A 1%
Wind speed 0.2m/s
Environmental requirement 0~40℃, relative humidity≤80%, in the place of no obvious vibration and corrosive gas
Equipment size & weight 1100mm* 550mm* 1100mm, 180KGS
 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Sinuo Testing Equipment Co. , Limited
Location Room 101, 1st Floor, No. 6, Third Street, Pingshan Industrial Zone 511495, Shibi Street, Panyu District, Guangzhou, China
Contact Person Zoe Zou

Request A Quote

Please check your email address.
Your message must be at least 20 characters.