Preirus Troubleshoot Ultrasound Machine Repair Hitachi Aloka
Hitachi Aloka HI VISION Preirus troubleshoot color image interference
Fault analysis and Troubleshooting
The color ultrasound diagnostic apparatus has a hot wheel interference during use, which seriously affects the diagnosis of the image.
According to the fault phenomenon, each probe of the machine in use has the same interference, and the possibility of failure of the color ultrasonic diagnostic apparatus is preliminarily ruled out.
According to the safe use specifications of the color ultrasonic diagnostic equipment, the grounding end of the machine shell and the machine power plug should be grounded. The loop formed by improper grounding can also cause electromagnetic interference; then, use a power measuring instrument to measure the grounding resistance. 2 ~ 5Ω, but after checking that the machine is not installed with a ground wire; after reinstalling the ground wire, the resistance value of the test ground wire is 3Ω, which meets the standard; the power-on test, the interference disappears, and the fault is eliminated.

Ultrasound parts that we can repair:
| Repair List of Featured Ultrasonic Board & Assy | |||
| Brand | System | Description | Part Number |
| GE | Logiq P5 | CPU mainboard | 5168431 |
| GE | Logiq P6 | CPU mainboard | |
| GE | Logiq E9/vivid E9 | GFI2 | 5161631 |
| GE | Voluson E6/Voluson E8 | RSR | KTI301394-2/KTI196357 |
| GE | Voluson E6/Voluson E8 | RFI/ RFI21b Board | KTI300614/KTI302197-6 |
| GE | Voluson S6/S8/P8 | BF64 | 5396937-2 |
| GE | Voluson S6/S8/P8 | RFS board | 5364098-2/5364098-3 |
| GE | Voluson S6/S10/P8 | BF192 board | 5357234 |
| Aloka | Alpha 5 | mainboard | EP493700DD |
| Aloka | Alpha 6 | RX Beam Former | EP555501AA |
| Aloka | Alpha 6 | CW board | EP555700 |
| Aloka | Alpha 6 | CPU mainboard | EP537000/EP563000 |
| Aloka | Alpha 6 | TX/RX | EP555300 |
| Aloka | Alpha 6/Alpha 7 | DOM | EP550000AA |
| Aloka | Alpha 10 | RX Beamformer board | EP531800CE |
| Aloka | Alpha 10 | wave generator Board | EP495000HH/EP495000HJ/EP495000FF |
| Hitachi | F37 | TX/RS | EP557400/EP557300 |
| Hitachi | F37 | mainboard | EP575700BC/EP560800 |
| Aloka | F75 | USP board | EP556600 |
| Aloka | F75 | Cell Board | EP558800 |
| Hitachi | HI VISION Avius | RDBF board | 7513907A/7513612 |
| Hitachi | Arietta 60 | CELL | EU-9160C |
| Hitachi | Arietta 70 | TX | EP572300AA |
| Hitachi | Arietta 70 | CELL | |
| Hitachi | Noblus | CELL | |
| Hitachi | Noblus | CONT-B | |
| Hitachi | Ascendus | CELL | 7352830A |
| Siemens | X150/G40 | BE | 10131804 |
| Siemens | X300 | BE 2.0/BE 3.0/BE 7.5 | 10131990 |
| Siemens | X700 | BE | 10136465 |
| Siemens | X700 | RC | 10136467 |
| Siemens | X700 | TI | 11014292 |
| Siemens | Antares | BE | 7472421 |
| Siemens | Antares | CB | 7288504 |
| Siemens | S2000 | RM300 | 10852163 |
| Siemens | S2000 high version/S3000 | RM301 | 10854749 |
| iU22/iE33 | EMB excelsior motherboard | 453561419431/453561382231/453561419431 | |
| iU22/iE33 | UMB Unified Motherboard | 453561254492/453561360251 | |
| HD15 | ACB acquisition control board | 453561197145/453561197144 | |
| HD15 | AIM board | 4.53561E+11 | |
| CX50 | control panel | 4.53562E+11 | |
| CX50 | mainboard | 453561368033/453561622163/453561495832 | |
| Affiniti 70 | ACQ acquisition control board | ||
| EPIQ 5/EPIQ 7 | ACQ Acquisition Module board | 453561734844/453561704246 | |
| Esaote | Mylab Series | CPU mainboard | 9501361000 |
| Toshiba | Aplio 300/400/500 | mainboard | PM30-38696 |
| Toshiba | Aplio 500 | RX | ywm2056*a |
| Toshiba | Xario 200 | TX | PM30-32732-1 |
| Toshiba | Xario 200 | RX | PM30-32733-1 |
| Remarks | |||
| 1.The repair cycle is within 3-5 days. | |||
| 2.Warranty after fixed:60 days. | |||
| 3. Kindly send us the estimated main faults (better to have image) for our engineer to prejudge. | |||
| 4.Package price is available if repair 3 pcs or more at one time. | |||
| 5.Extra discount for VIP partners. | |||
| 6.You could have different part No.under same description. | |||
The solution of dealing with ultrasonic image interference
1. Probe interference: replace the probe
2. Equipment interference: replace the power supply or internal boards
3. External interference:
(1) Workstation interference: it is recommended to add several magnetic rings to the video cable, if it is not work, replace the ATX power supply, or add a filter to the host power input.
(2) Power input interference: add several magnetic rings to the power line.
If the above methods do not work, just try the method of eliminating the equipotentiality
, adding a shielding cloth on the bed, and connect a wire to the ground strap of the machine.