GE Dash1800 Dash2500 Patient Monitor Parts Parameter Module DAS Board Assembly Masima And Necllor

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 3-5 working days
Brand: GE
Product Description

GE Dash1800 Dash2500 patient monitor parameter DAS module DAS Board assembly Masima and Necllor

 

 

 

 

► Brief Introduction

The DAS assembly is a field replaceable unit only. There are NO field repairs or adjustments for the DAS assembly. Please Do NOT open the DAS assembly as this breaks the isolation barrier which may result in patient death or serious injury.

 

 

► Replace DAS Board Assembly

1. Remove the recorder assembly first.

2. Remove the four screws from the rear case assembly and take out the chassis assembly.

3. Pull out the DAS assembly and replace with the new DAS assembly. There are two DAS assemblies to select: DAS assembly with Nellcorrr ® SpO2 or DAS assembly with Masima® SpO2.

 

 

Functional tests:

1. SpO2: The SpO2 sensor used depends on the screen configuration.

Nellcorrr® —SpO2 screens use a two-part assembly: DS-100A and NELL1 GE.

Masima SET screens® --SpO2 use a set of interface cable and sensor.

Connect the cables before connecting them to the screen. The SpO2 reading is displayed within moments of connecting the sensor to either the SpO2 simulator or your finger.

2. Blood pressure: A blood pressure test is performed by connecting the hose and belt attached together, and then connecting them to the NIBP connector on the left side of the screen. Press the NIBP Go/Stop hard key in the foreground to start the NIBP cycle.

3. ECG: Electrocardiogram monitoring uses the formation of 3 electrodes or 5 electrodes.

3-Lead ECG connection - Connect the ECG lead connector to the ECG box cable before connecting it to the screen. The simplest way to test the functions of an electrocardiogram circuit is through the use of an electrocardiogram simulator.

5-Lead ECG connection - Connect the ECG lead connector to the ECG box cable before connecting it to the screen. The simplest way to test the functions of an electrocardiogram circuit is through the use of an electrocardiogram simulator.

4. Temperature: Connect the temperature probe with a device

Interview connection. The predictive temperature begins as soon as the probe is removed from the clipboard. Replace the sensor after the TEMP cycle is finished.

 

Nellcorrr SpO2

Range 20 to 250 bpm
Accuracy and tolerance 20 to 250 bpm ± 3 digits
Low perfusion 20 to 250 bpm ± 3 digits

 

Measurement range SpO2 1 to 100%
Pulse rate 20 to 250 bpm

 

 

Masima SpO2

 

Range 25 to 240 bpm
Accuracy and motion tolerance Without motion 25 to 240 bpm ± 3 digits
Accuracy and motion tolerance With motion normal PHiliphhilipysiologic range, 25 to 240 bpm ± 5 digits
Measurement range SpO2 1 to 100%
Pulse rate 25 to 240 bpm

 

Please kindly contact us, if you are interested in GE Dash1800 Dash2500 patient monitor parameter DAS module DAS Board assembly Masima and Necllor, Thank you!

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.