Medical Equipment Parts Goldway UT-4000B UT-4000A UT-4000C UT4000Fpro UT-6000A Patient Monitor SpO2 board

 

 

 

► Brief Introduction

This SpO2 board is work for Goldway UT-4000B UT-4000A UT-4000C UT4000Fpro UT-6000A Patient Monitor.

We are trader specialized in the research, development and production of medical equipment parts. We are located in Guangzhou, with convenient transportation access. After many years of marketing and developing, we are quiet proficient in medical machines technology. All of our products comply with international quality standards and are greatly appreciated in variety of different markets throughout the world.If you are interested in any of our products or would you to discuss a custom order, please feel free to contact us. We are looking forward to forming successful business relationships with new clients around the world in the future.

 

 

► SpO2 Pulse Oximetry Test

  1. Power off the Oximetry simulator.
  2. Connect the Oximetry cable of the simulator to the SpO2 socket on the monitor.
  3. Power on the simulator, set simulator: SpO2 – 95%, Pulse Rate – 100
  4. Verify if SpO2 parameter area reading is between 92 – 99%, and if the waveform in the waveform channel is normal, and if the pulse rate reading is between 96 to 104.
  5. Power off the simulator and unplug the oximetry cable. Pulse oximetry test is complete.

 

► Specifications

Measure Range 50 ~ 100%
SpO2 Averaging 8 second average
SpO2 Accuracy ±2% (90 ~ 100%), ±3% (50 ~ 89%)
Pulse Rate Range 30 ~ 250 bpm
Pulse Rate Averaging 8 beat average
Pulse Rate Accuracy ±1% ( 30 ~ 100 bpm)
Sensor Types Finger, Universal “Y”, wrap probes
Pulse Rate Display Digital
Refurbish time about 2 seconds
Wavelength Infrared :940 nm
Red :660 nm
Energy Infrared ≤ 22.5 Mw
Red ≤ 30 mW

 

 

No SpO2 reading

PHiliphhilipenomenon:

In the process of monitoring, there is no SpO2 waveform and data.

 

Inspection method:

Check the finger probe to see if there is red light, if the arm is being pressed, and if the room temperature is too low.

 

Solution:

If there is no red light in the finger probe, check the extension cable and connecting part. When room temperature is very low, do not expose patient’s arm outside , which will affect the measurement results. Do not measure blood pressure and SpO2 on the same arm. Pressing the arm will affect SpO2 measurement.

If SpO2 waveform channel shows “No signal received”, it indicates that there is a problem in the communication between SpO2 module and the main system. If, after turning the monitor off and on again, the message still remains, change the SpO2 board.

 

 

Thank you for your time and interest in Medical Equipment Parts Goldway UT-4000B UT-4000A UT-4000C UT4000Fpro UT-6000A Patient Monitor SpO2 board

If you want to explore more details please let us know. 

Any inquiry will be highly appreciated and thanks for your support. 