P/N 1076374(1058272) PHiliphhilip V60 Ventilator Battery 14.4V 11.0Ah 163Wh LOT M91484-P1

 

 

Install an optional battery

Take the following precautions with the battery to reduce the risk of fire, explosion, leakage, or other hazard.

- Do not disassemble, open, drop, crush, bend, deform, insert foreign objects, puncture or attempt to crush the battery. Modify or revise it. Soaking or exposure to water or other liquids. Exposed to fire and overheated (including welding machines). Or put in the microwave.

- Replace the battery only with other batteries specified by the manufacturer.

- Follow all instructions for proper battery use.

- Do not short-circuit the battery or allow metal or conductive objects to come into contact with the battery connector housing.

-Use battery only with Respironics V60 ventilator

 

Specification

Technology Li - Ion
Voltage 14.4 v
Weight (kg) 1.59kg
Part No 1076374
Capacitance 11.00 A/h
Manufacturer RESPIRONICS
Used model Respirator V60
Maximum system current draw 11 A
Charge voltage +16.9 V maximum
Operating time 360 minutes under normal conditions

 

 

Install the battery as follows

 

1. Turn off and unplug the ventilator

NOTE: Failure to properly turn off the ventilator before installing the battery may result in false alarms after power on.

 

2. Turn captive fasteners 1/4 turn and loosen to remove side panel.

3. Using a 3mm hex key, remove the two screws to remove the battery holder.

4. Holding the battery with the vents facing up and the PHiliphhilip logo facing out, route the battery cable through the battery holder.

Position the battery and place it in the battery compartment.

Pinch the end of the battery connector and insert it so that it locks into place.

5. Replace the two screws and reinstall the battery bracket. Reinstall the side panel and secure the fasteners ¼ turn clockwise.

6. Make sure the battery is installed correctly by plugging the ventilator into an AC outlet and verifying that the yellow battery (charged) LED on the front panel is flashing. Blinking LED indicates battery is charging

 

 

Warning: Never attempt to disconnect or connect the battery while it is in use. NOTE: If the Check Vent or Vent Not Working alarm appears after battery installation while checking ventilator operation, stop using the ventilator immediately and contact PHiliphhilip. If AC power is disconnected and no battery is installed, or if the battery is completely discharged, a Ventilation Inoperative alarm will display.

NOTE: New batteries must be charged for at least 5 hours before being used.

 

 

FAQ:

 

1.Do you have English user manual for every item?
Most products are equipped with English instruction manual and operation guide.
 
2.Are your accessories safe for pediatric medical equipment?
All materials meet medical safety standards, suitable for adult and pediatric clinical equipment.
 
3.Do you supply veterinary medical device accessories?
Yes, we also provide compatible spare parts for veterinary medical equipment.
 
4.Can you customize special length cables?
Yes, we can customize different length and interface medical data cables.
 
5.Do you have various connector types for medical cables?
We supply full range of original and compatible connectors for medical cable replacement.

 

 

If you're interested in P/N 1076374(1058272) PHiliphhilip V60 Ventilator Battery 14.4V 11.0Ah 163Wh LOT M91484-P1,feel free to contact us .Thank you.

 

 

 


90 Day Warranty                             SGS ceryification

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.