EDAN F2 F3 F6 F9 TOCO Transducer MS3-31527(B) 02.01.210259013

Price Negotiable
Price: Negotiable
MOQ: 1
Delivery Time: 3-5 working days
Brand: Edan
Product Description

EDAN F2 F3 F6 F9 TOCO Transducer Blue label MS3-31527(B) TPU REF 02.01.210259 MPN02.01.210259012 (6)06944413805507 (21)H18900830290

 

 


Brief Introduction

Place the TOCO transducer on the patient‟s abdomen, which is flat and approximately 3 cm away from the fundus, e.g. slightly above the umbilicus on the left or on the right. The position should be different for different purposes: place the transducer close to the fetal buttocks for NST, and place it on fetal back in delivery. Wrap the abdomen with the belt over the transducer. Fix the transducer by pushing its buckle through the overlapping section of the belt. Make sure the belt fits the patient snugly but comfortably

Degree of protection against harmful ingress of water of the US transducer and TOCO transducer is IPX8. They can be continuously immersed in water to a depth of 1.1 meter for 24 hours and remain safe, but they are not allowed to be immersed in organic solvent, such as ethanol.

1. TOCOS Transducer Sensor (Blue Labeled)

2. Transducer Cable

3. Transducer Connector

 

 

► Testing TOCO Transducers

1. Switch on the monitor.

2. Connect the TOCO transducer to the fetal monitor.

3. Gently press the center of the transducer.

4. Check that the value on the display shows this change in pressure

5. Lay the transducer face up on a hard, flat surface for a few seconds.

6. Press the AUTO ZERO key to zero the TOCO displays 10.

7. Find a flat thing, such as a coin, whose diameter is less than 2.7cm.

8. Turn the transducer over so that the transducer is resting on the coin. The TOCO numeric should read between 35 and 50.

If a TOCO transducer fails the test, repeat this test with another transducer. If the second one passes the test, defect of the first transducer is confirmed. Replace it with a good one.

 

 

► Specifications

TOCO Range: 0% ~ 100%
Non-linear Error: < +10%
Resolution: 1%
Zero Mode: Automatic / Manual
Dielectric Strength > 4000Vrms

 

 

► Quality Guarantee

"Quality assurance" is our eternal pursuit and promise. We have a professional division of labor, a highly cooperative team, and a strong technical force in medical device development. Our expert team provides fast and reliable solutions for the sales and maintenance of patient monitors, defibrillators, ECG/ECG machines and other accessories or repair parts. We have always focused on producing quality products and providing first-class customer service. It is important to note that there are brand new and second-hand products for sale. Please distinguish them carefully when placing an order. Otherwise it will cause unnecessary trouble. If you have any questions about product quality, please contact us. We will get back to you as soon as possible.

 

 

Thank you for your time! If you have other questions about our EDAN F2 F3 F6 F9 TOCO Transducer MS3-31527(B) 02.01.210259013​ , please feel free to contact us. We are always ready to serve you.

 

 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company China World Technology Medical Equipment Service Group
Location TianHe District,GuangZhou,GuangDong
Contact Person Kiara

Request A Quote

Please check your email address.
Your message must be at least 20 characters.