408L Environmental Testing Temperature Humidity Chamber -40°C to +180°C Range with 20-98% RH Control

Price Negotiable
Price: Customized
MOQ: 1
Delivery Time: 45 days
Brand: Sinuo
Product Description
Compliance Standards & Certifications
  • IEC60068-2-1:2007 "Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT"
  • IEC60068-2-2:2007 "Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT"
  • IEC60068-2-30:2005 "Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT"
  • IEC60068-2-78:2012 "Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT"
Equipment Overview

This environmental testing chamber performs cold, heat, humidity, and dryness testing for raw materials and devices including electronics, electricians, electrical appliances, and plastics. The equipment verifies product reliability during storage and transportation under alternating high/low temperature and humidity conditions.

Technical Specifications
Parameter Specification
Control System PLC and touch screen intelligent control
Chamber Size (Internal) 800×600×850 mm (408L capacity)
External Dimensions 1350×1150×1950 mm
Temperature Range -40°C to +180°C
Humidity Range 20% to 98% RH
Temperature Fluctuation ±0.5°C
Humidity Fluctuation ≤2.5%
Temperature Deviation ≤±2.0°C
Humidity Deviation ≤±3%
Heating Rate ≥2-3°C/min
Cooling Rate ≥1°C/min (average no load)
Installation Requirements
Site Conditions

Flat ground with good ventilation, free from flammable, explosive, corrosive gases and dust. No strong electromagnetic radiation sources nearby. Drainage floor drain required within 2 meters of refrigeration unit. Ground bearing capacity: ≥800kg/m². Minimum maintenance space: front 600mm, sides 1000mm, rear 100mm, top 600mm.

Environmental testing chamber installation space requirements diagram
Water Source Requirements
  • Purified water must be used
  • Tap water requires treatment
Environmental Conditions
  • Temperature: 5°C to 35°C
  • Relative humidity: ≤85%
  • Air pressure: 86kPa to 106kPa
Power Requirements
  • Power supply: AC380V three-phase four-wire + protective ground wire
  • Voltage fluctuation: AC (380±10)V
  • Frequency fluctuation: (50±0.5)Hz
  • Ground resistance:
  • Power capacity: ~11kW
  • Maximum current: 23A
  • Power switch: 32A molded case leakage protector
Storage Conditions

When not in operation, maintain ambient temperature between 0°C and +45°C.

Product Images
Front view of 408L environmental testing temperature and humidity chamber Control panel and interior view of temperature humidity test chamber

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Sinuo Testing Equipment Co. , Limited
Location Room 101, 1st Floor, No. 6, Third Street, Pingshan Industrial Zone, Shibi Street, Panyu District, Guangzhou, China
Contact Person Zoe Zou

Request A Quote

Please check your email address.
Your message must be at least 20 characters.