Hematology Analyzer Reagents Abbott Analyzer CD-3000 Medical Equipment Analyzer Reagents

 

Intended Use

To be used for blood cell counting, sizing, and WBC differentiation, hemoglobin determination on ABBOTT CD3000,3500,3700 hematology analyzers. This product is for In Vitro Diagnostic Use only.

APPLICABLE INSTUMENT

For use with CD3000,3500,3700 hematology analyzers.

PRECAUTIONS
  • Re-calibration is recommended when changing the origin of the reagents.
  • It is not recommended using different brands of solutions together on one unit of analyzer.
  • Do not ingest. Avoid contact with skin and eyes.
  • Skin contact: Flush with plenty of water.
  • Eyes contact: Flush with plenty of water. Obtain immediate medical attention.
  • This kit is for professionals and in vitro diagnostic use only.
  • Never put distilled water on the analyser (except clean for shipping menu), otherwise the pneumatic unit can be damaged seriously.
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Specification

L-99231-01 Diluent CD5 20L
L-99311-01 Sheath CDS 10L
L-99321-01 Detergent CD5 20L
33-40-E001 EZ-Cleaner 50ml×2

 

KIT COMPOSITION

Diluent CD5 NaCl≤9.4%, Buffer≤1.6%, Anti-Microbial Agent≤1.0%
Detergent CD5

Sodium Sulfate<1.5%, Sodium Chloride<0.6%

Sheath CDS Sodium Phosphate, Diabasic<0.3%
EZ-Cleaner Subtilisin<5.0%
 

PROCEDURES

  • Using the function keys press the following menus:

MAIN-SPECIAL PROTOCOLS-REAGENT RESERVOIRE-EMTY DILUENT

  • Remove the reagent inlet tube from the Diluent.
  • Wipe the tube’s surface with a lent free tissue, and then rinse with distilled water.
  • Put the reagent tubes into the boxes and/or bottles of Labnovation reagents.
  • Using the function keys press the following menu:

FILL DILUENT

This step will fill the reagent system with Labnovation reagent.

  • Repeat steps 1-5 with Lyse and Sheath.
  • Run at least 5 backgrounds. Check for the acceptable values.
  • Run all three levels of a dedicated commercial control. Check for the acceptable values.
  • Recalibrate the unit if necessary. (MAIN-CALIBRATION-ENTER-FACTOR)

 

Advantage
  • One source for all brands of analyzer 
  • Reagents for both 3-parts and 5-parts
  • Ready-to-use final products and bulk OEM reagents
  • ISQ13485:2003 and ISO9001:2008 certified
  • CE marked, FDA cleared
  • Well-established brand used by thousands of lab in 95 countries
  • High quality, high performance and high fidelity of correlations with original brand
 
 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Labnovation Technologies, Inc.
Location 101 and 5th Floor, Building 1, No. 68,18th Road, Guangming Hi-Tech Park, Tangjia Community, Fenghuang Street, Guangming District, Shenzhen 518107 ,China
Contact Person Aimee li

Request A Quote

Please check your email address.
Your message must be at least 20 characters.