Real Time PCR Mycoplasma Genitalium MG STD STI Test Kit Lyophilized 48 Tests/Kit

Price Negotiable
Price: USD
MOQ: We can produce liquid and lyophilized kits
Delivery Time: Depending on order quantity
Brand: Biokey
Product Description
Mycoplasma genitalium (MG) real-time PCR detection kit (lyophilized)
--48 tests/kit

 
Intended use:
 

This kit, adopting Real-Time PCR-Fluorescence technology, is used for detection of MG. It can be used for the laboratory diagnosis and monitoring of MG infection. The test results are only for clinical reference, not for case confirmation or exclusion.

 

Principle:
 

The specific primers and specific fluorescent probes are designed for the nucleic acid conserved region of MG. The PCR reaction solution and real-time

 

MG product lyophilization pacakage:

 

 
Main components:

 

NO. Components

BIK-QL-H0019S

1

PCR Enzyme Mix(Lyophilized)

48 tests/kit

2 Positive Control 100 μL/tube
3 Negative Control 100 μL/tube
4 RNase-free water 1mL/tube
5

Paraffin oil

1.5mL/tube

 
Note: Do not mix the components from different batches for detection. The Positive Control of MG and internal control
were constructed artificially, and they were not infectious.

 
Specimen requirements:
 

1. Specimen type

Urine or urethral swabs for men, and female urine, cervical swabs, or vaginal swab samples

2. Specimen Preservation

The specimen can be stored for 3 months at -20±5℃ and for a long time below -70℃

 

 

PCR Amplification condition:

 
PCR amplification according to the following procedures:
 

Step Cycle number Temperature Time
1 1 95℃ 3min
2 40 95℃ 10s
60℃ 30s collect fluorescent

 

Instrument detection channel selection: FAM and HEX/VIC.
 
Quality control:
 

  Ct value
Negative Control No Ct
Positive Control ≤30
IC of samples ≤38


Note: Negative Control, Positive Control and IC of samples must be performed correctly, otherwise the sample results is invalid.
 

Data analysis and interpretation:

 

The following sample results are possible:

  Ct value Result Analysis
1# No Ct negative
2# ≤38 Positive
3# 38~40 Re-test; if it is still 38~40, report as 2#


 Operation details please check from IFU!

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Guangzhou BioKey Healthy Technology Co.Ltd
Location Add:No.2, Ruitai Rd, Guangzhou Hi-Tech Industrial Development Zone, 510670, CHINA
Contact Person Lisa

Request A Quote

Please check your email address.
Your message must be at least 20 characters.