1.Description:

This instrument can simultaneously measure the heat storage coefficient, thermal conductivity, thermal diffusivity and specific heat of materials under different moisture contents in a single test. The device is simple, highly accurate and has a fast testing speed (about ten minutes for a single test).

2.Standard:

JGJT 12-2019 (replacing JGJ 51-2002) (Light Aggregate Concrete Technical Code), JG158-2004 "Adhesive Powder Polystyrene Granule External Wall External Insulation System", and Appendix B of JG/T283-2010 "Expanded Porous Microsphere Lightweight Mortar" regarding the determination of the heat storage coefficient, the requirements of JG/T511-2017 "Building Insulation Ceramic Insulation Board" Appendix A for the test method of the heat storage coefficient of foamed ceramic insulation board, and the determination method of the heat storage coefficient in GB/T26000-2010 "Expanded Porous Microsphere Insulation and Heat Insulation Mortar".

3.Working Principle

The instrument is based on the principle of non-steady heat conduction. It heats the experimental materials for a short period of time, causing changes in the temperature of the materials. Based on the characteristics of these changes, by solving the heat conduction differential equation, the heat storage coefficient, thermal conductivity, thermal diffusivity and specific heat of the experimental materials can be calculated.

4.Configuration:

Including one LRY-II mainframe, one set of software, one heater, one set of thermocouples, one insulated bucket, and one desktop computer (configured as per the contract).

5.Technical Parameter

Parameter Specification
Application Scope Measures thermal conductivity, thermal effusivity, thermal diffusivity, and specific heat capacity of homogeneous materials in dry or moist states:
• Forms: Plates, gels, powders, granules
• Thermal Effusivity Range: 0.1–30 W/m²·K
• Thermal Conductivity Range: 0.035–1.7 W/m·K
Sample Specifications Thin Sample:
20 × 20 × (1.5–3) cm (One piece)
Thick Samples:
20 × 20 × (4–10) cm (Two pieces)
Power Supply AC 220V, 50 Hz
Dimensions (L×W×H) Legacy Model (Split Type): 600 × 450 × 700 mm
New Model (Integrated): 340 × 700 × 560 mm
Operating Conditions • Ambient Temperature: 10–35°C
• Relative Humidity: ≤80% RH
• Temperature Stability: ≤±1.5°C
Measurement Accuracy ±5%
Data Interface PC connectivity with automatic testing, data recording, and print output
 

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company DONGGUAN LONROY EQUIPMENT CO LTD
Location A31, 11 Floor, No 1 Building, Minjian Jinrong Building, No 111 Guantai Road, Sanyuanli Community, Nancheng Street, Dongguan, Guangdong, China
Contact Person Kaitlyn Wang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.