280L Horizontal Cylindrical Autoclave HCPS-A280

Price Negotiable
Price: Available upon request
MOQ: 1 unit
Delivery Time: 10-15 working days
Brand: Senova
Product Description

280L Horizontal Cylindrical Autoclave HCPS-A280

Features:
  • Relies on gravity air removal for a more complete steam environment, enhancing sterilization reliability.
  • Automated process control regulates steam dynamically according to preset temperature parameters.
  • Versatile unit suitable for sterilizing a wide range of medical, laboratory, and clinical supplies.
  • Effective drying capability meets the post-sterilization processing needs of various materials.
  • Equipped with an automatic pressure relief valve (0.24 MPa) for operational safety.
  • Full safety package: auto-protection from excessive temperature/pressure and water shortage with alarm.
  • Secure door interlock prevents access until internal pressure is below 0.027 MPa.
  • Manufactured with a high-quality stainless steel chamber for durability and sanitation.
Specifications
Technical data HCPS-A280
Chamber volume 280L
Chamber size φ600*1000mm
Working pressure 0.22MPa
Working temperature 134℃
Adjustment of temperature 105℃ - 134℃
Sterilization time 0-60 min
Drying time 0-60 min
Heat average ≤±2℃
Power 12KW,380V/50Hz
Dimensions 1400*770*1780
Packing size 1680*920*2100
G.W/N.W(Kg) 465/365
Introduction about autoclave

An autoclave is a device designed to achieve sterility through the application of pressurized steam. The principle is based on the fact that moist heat is more effective at killing microbes than dry heat at the same temperature. By containing steam under pressure (e.g., 15-30 psi), temperatures of 121°C to 134°C are readily achieved. The standardized cycle ensures thorough treatment: first, air is purged from the chamber; second, steam penetrates the load for a validated time; third, steam is exhausted and items are dried. Its applications are fundamental to safe practice in medicine, dentistry, and laboratory science. It processes reusable instruments, prepares sterile culture media, and treats infectious waste.

In industry, it is used for processes requiring heat and pressure, such as bonding composite materials. The autoclave is the gold standard because it provides a reliable, repeatable, and monitorable process. Biological and chemical indicators can verify its efficacy. Steam's excellent heat transfer properties ensure that even complex geometries are uniformly heated, providing a level of assurance that alternative methods often cannot match for the same breadth of materials.

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company SENOVA BIOTECH (SHANGHAI) CO., LTD.
Location No. 22, Lane 295, Jigao Road, MinHang District, Shanghai 201107, China
Contact Person Ricky Chang

Request A Quote

Please check your email address.
Your message must be at least 20 characters.