UPS / Medical Instruments Sealed Lead Acid Battery 6V Rechargeable Battery
UPS / Medical Instruments Sealed Lead Acid Battery 6V Rechargeable Battery
Specifications:
| Nominal Voltage | 6V | |
| Nominal Capacity (C20) | 3.4Ah (5.25V,at 25℃) | |
| Dimension | Length | 134mm |
| Width | 34mm | |
| Height | 60mm | |
| Total Height | 67mm | |
| Weight | Approx.0.69 kg | |
| Terminal | FO-01 | |
| Internal Resistance | Approx.25 mΩ (25℃) | |
| Max Charging Current | 1.02A | |
Description:
● Designed life for floating charge is 3~5 years (20℃);
● AGM valve regulated sealing technology;
● High strength ABS material for container & lid;
● Wide temperature scope of application (-15~45℃);
● Best temperature of application (20±5℃);
● No leaking, safe and reliable;
● High sealed reaction efficiency, little loss of water, no need to add distilled water or electrolyte, simple to use and maintain;
● Low self-discharge rate.
Applications:
● UPS & EPS system
● Telecommunication
● Banks & Financial Centre
● Hospital& Testing Laboratories
● Emergency Power Systerm
● Alarm & Security system
● High Power Backup Supply
Charge
| Using Mode | Charging Voltage | Temperature Compensation | Max Charging Current |
| Standby Use | 2.275±0.025V/cell (25℃) | -3.3mV/℃/cell | 1.02A |
| Cyclic Use | 2.45±0.05V/cell (25℃) | -5mV/℃/cell |
TECHNICAL FEATURES:
• Float Service Life
CB, HR and SE Series batteries have an expected life span of 3 to 5 years in float service applications.
• Self Discharge - Shelf Life
The self discharge rate at room temperature is approximately 3% of rated capacity per month.
• Operating Temperature
CGB Batteries may be operated over a broad range of ambient temperatures.
• Deep Discharge Recovery
The batteries recover their capacities even after repeated deep discharges.