Oxidation Stability Tester for Inhibited Mineral Oil ASTM D943
This Inhibited Mineral Oil Oxidation Stability Tester is suitable for determining the oxidative stability of unused mineral insulating oils with inhibitors, and can also be used for the tendency of unused mineral insulating oils to form acids and precipitates before and after the induction period. Mechanism: The sample was kept at 120°C in the presence of the copper catalyst coil, while a constant flow of oxygen was passed through, and the time required for the volatile acid generated by the oxidation of the sample to increase to a neutralization value equivalent to 0.28 gKOH/g was measured. The oxidative stability of the sample is represented by the induction period (h).
- Metal bath heating, safe and environmentally friendly.
- Microcomputer temperature controller with digital display, accuracy ±0.1℃, equipped with PT100 sensor.
- Gas flow controlled by precision flow meter.
- Capable of running 6 tests simultaneously, divided into two groups with 3 samples each.
- Digital display of operating time.
- Glass test tubes meet standard requirements.
- High-power solid-state relay for temperature rise control.
- Over-temperature alarm function.

|
Applicable standards
|
ASTM D943
|
|
Heating mode
|
Electric heating rod
|
|
Temperature Control Mode
|
Imported digital display PID temperature controller
|
|
Temperature Control Accuracy
|
±0.1℃
|
|
Temperature Control Range
|
Ambient temperature ~ 120℃
|
|
Working units
|
6 Units
|
|
Timing Mode
|
Digital display timer
|
|
Flow Range
|
0 ~ 200 mL/h
|
|
Flow Control
|
Precision flow meter
|
|
Total Power
|
About 2.5KW
|
|
Dimension
|
630mm*710mm*930mm
|
|
Net Weight
|
50 kg
|
|
Power supply
|
AC 220V ,50Hz
|

Get in Touch