Easy Installation ISO9239-1 Building Material Flooring Radiant Panel Tester
Introduction:
-
This Flooring Radiant Panel Tester is the cornerstone of the flooring materials reaction to fire tests in the Euroclass system.
-
The test measures the spread of flame of the sample, under thermal exposure, using a radiant heat ignition source. The thermal exposure is that of a fire growing in the room of origin that through a door opening, radiates onto the floorings in an adjacent room or corridor.
-
This Flooring Radiant Panel Tester an also be used for tests to standards ASTM E648 and DIN 4102; Part 14.
Features:
-
The resulting test chamber is insulated with calcium silicate insulation board and is provided with a temperature resistant observation window.
-
The stainless steel sample support assembly is mounted on a sliding platform to allow safe and easy loading of the test sample.
-
The radiant heat is applied by means of a gas-fuelled panel, inclined at 30º, and directed at a horizontally mounted floor covering system specimen.
-
The radiant panel generates a radiant energy flux distribution ranging from a nominal maximum of 10.9 kW/m2 to a minimum of 1.1 kW/m2.
-
Dummy calibration specimen with holder, calibrated heat flux meter and mounting on the Flooring Radiant Panel Tester .
-
The distance burned until flame-out is reached and converted, by calibration, into an equivalent critical radiant flux, in kW/m2.
-
Heat Flux Meter: Range 0 ~ 50 kw/m2; Surface emissivity: ε=0.95± 0.05
-
A smoke measuring system, according to DIN 50055, is mounted on a separate frame at the exhaust stack.
-
Control Rack for convenience in use, allowing observation of the apparatus and controls during equipment set-up and calibration.
-
Automatic ignition of the radiant panel and safety cut-out.
-
Automatic moving T type ignition burner is supplied with Flooring Radiant Panel Tester .
-
Stainless steel hood with smoke measurement ports.
Parameters:
|
1 |
Dimension |
1900mm (W) x 750mm x (H) x 2120mm (D) |
|
2 |
Weight |
240kg |
|
3 |
Power supply |
110V AC 60Hz / 230V AC 50Hz, 10A |
|
4 |
Ambient temperature |
10 ~ 35 Centigrade |
|
5 |
Burning gas |
Propane |
|
6 |
Auxiliary gas |
Air compressor gas |
Get in Touch