Tunkia Co., Ltd.
                                                                                                           
Verified Supplier
28 Years
Since 1998
Menu

TD1350 Electrical Measuring Instruments Ammeter AC And DC High Precision

Price Negotiable
Price: negotiation
MOQ: negotiation
Delivery Time: Negotiable
Brand: TUNKIA
Product Description
TD1350 High-Precision AC/DC Ammeter
 

Introduction

  • TD1350, a reference AC/DC ammeter with an ultra-wide current measurement range. The accuracy is up to class 0.005% and 0.01%, and the measurement accuracy of DC current above 1 A and AC current is much better than 8 digital meters.

  • Instead of 8 digital meters, TD1350 is applied to measure aperiodic signals (e.g. ripple) or complex waveform signals and calibrate AC/DC current sources (e.g. Fluke 5500A/5520A) and multimeter calibrators which the accuracy classes are 0.02% and below.

Features

  • Wide range of direct current measurement: a pair of terminals can be used to measure the DC current of 10 A ~ 120A, which solves the top problem of automatic range switching of ultra-wide range DC current.

  • Current expansion measurement: it is equipped with a special measurement interface for the sensor, which can be connected with a through current sensor to expand the measurement range to 500 A, 1 kA, etc.

  • High Reliability: Not easy to damage when overloaded.

  • Other Measurements: Ripple, harmonic, distortion, waveform monitoring, etc.

  • Multiple measurement modes: DC, AC, DC + AC three modes, fundamental frequency: 10 Hz ~ 1 kHz, and measure periodic or aperiodic signals.

  • Excellent Human-computer Interaction: LCD touch screen, more intuitive of value display and easier operation. it is equipped with RS232, USB, RJ45 interfaces, which is convenient to set up a full-automatic measurement system.

Specification

 

DC Current Measurement DCI

Range Resolution 0.01% 0.005%

Accuracy

ppm*RD +ppm*RG or A

Temperature Coefficient

ppm*RD /℃

Accuracy

ppm*RD +ppm*RG or A

Temperature Coefficient

ppm*RD /℃

24 Hours 90 Days 1 Year 24 Hours 90 Days 1 Year
100 μA 0.1 nA 60 + 50 40 + 50 60 + 50 5 + 1 n 30 + 50 40 + 50 60 + 50 3 + 0.5 n
300 μA 0.1 nA 60 + 50 40 + 50 60 + 50 5 + 2 n 30 + 50 40 + 50 60 + 50 3 + 1.5 n
1 mA 1 nA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
3 mA 1 nA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
10 mA 10 nA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
30 mA 10 nA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
100 mA 100 nA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
300 mA 100 nA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
1 A 1 μA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
3 A 1 μA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
10 A 10 μA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
30 A 10 μA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
100 A 100 μA 30 + 30 40 + 30 60 + 40 5 + 1.5 15 + 15 20 + 15 30 + 20 2.5 + 1
  • Measuring Range: ± (10 μA ~ 120 A)

  • Range Switching: Automatic or Manual

  • Ripple measurement: 1 Hz ~ 10 kHz

  • Accuracy: 0.02%*RG, effective value

 

AC Current Measurement

Range Resolution 0.01% 0.005%

Accuracy (k = 2)

ppm*RD +ppm*RG

Temperature Coefficient

ppm*RD /℃

Accuracy (k = 2)

ppm*RD +ppm*RG

Temperature Coefficient

ppm*RD /℃

40Hz≤F≤200Hz 200Hz≤F≤10kHz 40Hz≤F≤200Hz 200Hz≤F≤10kHz
100 μA 0.1 nA 250 + 150 300 + 200 5 + 5 200 + 100 250 + 150 3 + 5
300 μA 0.1 nA 250 + 150 300 + 200 5 + 5 200 + 100 250 + 150 3 + 5
1 mA 1 nA 250 + 150 300 + 200 3 + 2 200 + 100 250 + 150 2.5 + 1
3 mA 1 nA 100 + 60 120 + 80 3 + 2 60 + 40 100 + 50 2.5 + 1
10 mA 10 nA 100 + 60 120 + 80 3 + 2 60 + 40 100 + 50 2.5 + 1
30 mA 10 nA 100 + 60 120 + 80 3 + 2 60 + 40 100 + 50 2.5 + 1
100 mA 100 nA 100 + 60 120 + 80 3 + 2 60 + 40 100 + 50 2.5 + 1
300 mA 100 nA 100 + 60 120 + 80 3 + 2 60 + 40 100 + 50 2.5 + 1
1 A 1 μA 250 + 150 300 + 200 3 + 2 200 + 100 250 + 150 2.5 + 1
3 A 1 μA 250 + 150 300 + 200 3 + 2 200 + 100 250 + 150 2.5 + 1
10 A 10 μA 250 + 150 300 + 200 3 + 2 200 + 100 250 + 150 2.5 + 1
30 A 10 μA 250 + 150 300 + 200 3 + 2 200 + 100 250 + 150 2.5 + 1
100 A 100 μA 250 + 150 300 + 200 3 + 2 200 + 100 250 + 150 2.5 + 1
  • Measuring Range: 10 μA ~ 120 A

  • Range Switching: Automatic or Manual; Display: 7 digits

  • Ripple measurement: 10 Hz ~ 1 kHz

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company Tunkia Co., Ltd.

Request A Quote

Please check your email address.
Your message must be at least 20 characters.