DC Magnetic Testing System TD8220 For Soft Magnetic Material
Product Introduction
The TD8220 is a intelligent instruments applicable for measuring various DC magnetic properties of soft magnetic materias. It meets national standard GB/T 13012-2008GB/T 3656-2008GB/T 13888-2009GB/T 35690-2017, and international standards IEC 60404-4, IEC 60404-7IEC 60404-15.
Applications
-
Measuring materialsSoft ferrite, electrical pure iron, silicon steel, NiFe (permalloy), CoFe alloy, amorphous and nanocrystalline alloys, metal magnetic powder core, etc.
-
Sample shape: Ring or open-circuit specimens measurement..
-
Supports simulation impact method and scanning method for measurement.
-
Measurable DC parametersBm, Br ,Hc, i, m.
-
Equipped with PC software. Draw magnetization curve Hysteresis loop and permeability curve .
Technical Specifications
Magnetic Parametes
|
Sample |
Method |
Parameters |
Uncertainty ( k = 2 ) |
Repeatability |
|
Ring |
impactmethod |
Bs |
1.0% |
0.3% |
|
Br |
1.0% |
0.3% |
||
|
Hc |
1.0% |
0.5% |
||
|
i |
4.0% |
1.5% |
||
|
m |
2.0% |
1.0% |
||
|
Open-circuit |
Permeameter (impact method) |
Bs |
2.0% |
0.5% |
|
Permeameter/Electromagnet (scanning method) |
Bs |
2.0% |
0.5% |
|
|
Br |
2.0% |
0.5% |
||
|
Hc |
2.0% |
0.5% |
||
|
Solenoid |
Hc |
2.5% |
1.0% |
|
Output |
Current range |
0 ~ ± 6 A |
|
Maximum power |
120 W |
|
|
Resolution |
0.005%*RG ① |
|
|
Uncertainty |
0.2% |
|
|
Magnetic flux measurement |
Fluxmeter ranges |
500 μWb,5 mWb,25 mWb |
|
Drift |
± (0.05% * RG) / min or ± 1 μ WB / min, taking the maximum of two numbers |
|
|
Uncertainty |
0.3% + 5 μWb |
|
|
Magnetic field measurement |
range |
0~4000 mT |
|
Uncertainty |
≤2500 mT;± 0.5%*RG |
|
|
Feature |
Hall probe nonlinear correction function |
-
Test conditions: ( 23 ± 5 ) ° C , test after demagnetization of the ring sample.
-
The ring impact is measured by the simulated impact method, and the ratio of the outer diameter to the inner diameter should be less than 1.1.
|
Output |
Current range |
0 ~ ± 6 A |
|
Maximum power |
120 W |
|
|
Resolution |
0.005%*RG ① |
|
|
Uncertainty |
0.2% |
|
|
Magnetic flux measurement |
Fluxmeter ranges |
500 μWb,5 mWb,25 mWb |
|
Drift |
± (0.05% * RG) / min or ± 1 μ WB / min, taking the maximum of two numbers |
|
|
Uncertainty |
0.3% + 5 μWb |
|
|
Magnetic field measurement |
range |
0~4000 mT |
|
Uncertainty |
≤2500 mT;± 0.5%*RG |
|
|
Feature |
Hall probe nonlinear correction function |
PC Software
-
Support Windows system, easy to operate.
-
The test report contains complete curves, test results, test conditions and sample parameters.
-
Test results support direct printing, export JPG or Excel.
-
Automatically save measurement data, draw complete magnetic curves, and test conditions and sample parameters.
-
Get in Touch
Have questions about our products or want to discuss a custom order? Our team is ready to help you.